![Ohmeda Giraffe OmniBed Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/ohmeda/giraffe-omnibed/giraffe-omnibed_operators-manual_3546691051.webp)
A-1
6600-0342-000 08/08/01
A-1
Appendix
Specifications
Power Requirements
Accessory Outlets
10.5A @ 100v ~, 50/60 Hz
2A @ 100v ~, 50/60 Hz
9A @ 115v ~, 50/60 Hz
2A @ 115v ~, 50/60 Hz
4.5A @ 220v ~, 50/60 Hz
1A @ 220v ~, 50/60 Hz
4.5A @ 230v ~, 50/60 Hz
1A @ 230v ~, 50/60 Hz
4.5A @ 240v ~, 50/60 Hz
1A @ 240v ~, 50/60 Hz
Standards
Designed to meet requirements of
IEC 601-2-19 (Amendment 1)
IEC 601-2-21 (Amendment 1)
IEC 601-1
IEC 601-1-2
Operating Environment
Temperature
20 to 30°C
Humidity
10 to 95% Non-condensing relative humidity
Storage Conditions
Temperature
-25 to 60°C
Humidity
10 to 95% Non-condensing relative humidity
Pressure
50 to 106 kPa
User Control Settings
Patient control temperature
35-37.5°C in 0.1° increments
Air control temperature
20-39°C in 0.1 increments
Radiant heat power
0-100% in 5% increments
Humidity
Servo- % relative humidity
30-95% in 5% increments
Manual- increments
1-10
Alarms
High Air Temp
1.5°C over AST (air set temperature)
Low Air Temp
3.0°C under AST
Baby Hot
1.0°C* over BST (baby set temperature)
Baby Cold
1.0°C* under BST
Fan Failure
Failure of blower system
Air Temp >38°C
>38°C for AST <37°C
Air Temp >40°C
>40°C air temperature for AST >37°C
Air Probe Failure
Compartment air probe failure
Air Probe Disconnect
Disconnection of compartment air probe
Power failure
Power switch on but no power
System failure
Non-recoverable system failure
Add Water
Humidifier water level low
Scale
Weight Exceeds Maximum
Greater than 8kg load (visual only)
Scale Failure
Detectable system failure (visual only)
* Can be re-set on the Service screen to 0.5°C
Содержание Giraffe OmniBed
Страница 1: ...GiraffeTM OmniBedTM Operator s Manual...
Страница 24: ...1 20 6600 0342 000 08 08 01 1 20 General Information Notes...
Страница 28: ...2 4 6600 0342 000 08 08 01 2 4 Preoperative Checkout Procedure Notes...
Страница 48: ...4 8 6600 0342 000 08 08 01 4 8 Maintaining the OmniBed Notes...
Страница 50: ...Accessories 6600 0342 000 08 08 01 5 2 5 2...