EX Series Operating Manual
6-12. CAL-DEFAULT
▌This function is to restore to the state of calibration data that shipped from factory.
▌If you calibrate the unit in the condition that you have no previous knowledge of calibration or you
can not restore by using "5-10. CAL-RESTORE" function, you can return to factory default calibration
data. If you use this function, you must calibrate this unit at the official calibration center.
CAL-DEFAULT Operating
▌Power switch on in a pressing "FACTORY" key.
▌Power switch ON
In a pressing
▌When appearing "1. user-mem clear" message, release "FACTORY" key.
▌Enter to "10. CAL-DEFAULT" mode by using cursor key.
▌In other to restore, press the "FACTORY" key one more time.
LCD Display
done
Release
LCD Display
6-13. LOAD DEFAULT
▌This function is to return to default value that shipped the unit from factory. You can erase
setting and changing all value at the "Factory Mode"
▌Initializing contents.
done
> User Memory Clear
> Last State : Disable
> Auto Cursor Move : Disable
> Auto Key Lock : Disable > OVP Use : Enable
> OCP Use : Enable
> ADC Sampling : 20Hz
LOAD DEFAULT Operating
▌Power switch on in a pressing "FACTORY" key.
In a pressing
▌Power switch ON
▌When appearing "1. user-mem clear" message, release "FACTORY" key.
▌Enter to "11. LOAD DEFAULT" mode by using cursor key.
Release
▌In other to return to default value, press the "FACTORY" key
one more time.
LCD Display
≫ Related remote interface command
FACT:LOAD-DEF
done
Apply :
FACT:LOAD-DEF
Initialize all the factory setting value.
- 83 -
Содержание EX 10000
Страница 2: ......
Страница 104: ...EX Series Operating Manual 104...
Страница 105: ...EX Series Operating Manual 105...
Страница 107: ...EX Series Operating Manual 2U 2 6KW 5KW Product Outside Dimension 2U 2 6KW 5KW Product Rear Feature 107...
Страница 108: ...EX Series Operating Manual 4U 5 1KW 10KW Product Outside Dimension 4U 5 1KW 10KW Product Rear Feature 108...
Страница 113: ...EX Series Operating Manual Reliability Durability Life time Test System Solution 113...