EX Series Operating Manual
▌When occurs OVP trip이, "
OVP TRIPPED
" message appears on the
display
▌In other to clear tripped OVP, increase the OVP Level.
In other to increase OVP level, press the "Protection" key.
LCD Display
protect> 1.ovp
▌
In the
"
protect> 1.ovp
", press the "Protection" key one more time.
LCD Display
ovp set> 15.00v
▌Move the under bar cursor where you want to change the level.
▌
y
g
▌Readjust the OVP level by using the encoder knob.
▌Finished to change the level, press the "Protection" key.
LCD Display
CHANGED
After displayed above message, return to before state that tripped OVP.
≫ Related Remote Interface Command
VOLT:OVP {<voltage>}
VOLT:OVP?
VOL OV
I ?
VOLT:OVP:TRIP?
VOLT:OVP:CLE
Apply 1:
Method of checking tripped OVP or not and readjust the OVP level.
VOLT:OVP:TRIP? Return value "1"
Check the tripp OVP state
VOLT:OVP 32
Set the OVP level.
VOLT:OVP:CLE
Clear the OVP trip.
p
Apply 2:
Method of checking tripped OVP or not and decrease the OVP level.
VOLT:OVP:TRIP? Return value "1"
Check the tripp OVP state
VOLT 0
Set the voltage value to zero.
VOLT:OVP:CLE
Clear the OVP trip.
- 30 -
Содержание EX 10000
Страница 2: ......
Страница 104: ...EX Series Operating Manual 104...
Страница 105: ...EX Series Operating Manual 105...
Страница 107: ...EX Series Operating Manual 2U 2 6KW 5KW Product Outside Dimension 2U 2 6KW 5KW Product Rear Feature 107...
Страница 108: ...EX Series Operating Manual 4U 5 1KW 10KW Product Outside Dimension 4U 5 1KW 10KW Product Rear Feature 108...
Страница 113: ...EX Series Operating Manual Reliability Durability Life time Test System Solution 113...