Start the decon cycle by following the prompts on the NuTouch display screen.
The decon cycle will automatically go through the following phases:
HEAT UP
1 hour
Complete the chamber preparation and either schedule a start time or press the Start button to start the heat
up portion of the cycle.
•
Temperature display shows chamber temperature.
•
Text displays that the unit is currently in the Heating Up cycle
•
A timer tracks how long the unit has been heating thus far
DECONTAMINATION
9 hours
The unit automatically starts a timer for the correct length of cycle when the decontamination temperature is
reached.
•
Temperature display shows chamber temperature.
•
The Text displays that the Decontamination cycle is in progress
•
A timer tracks the remaining time in Decontamination
HUMIDITY REDUCTION
1 hour
The unit remains at the decontamination temperature and the air pump is turned on to reduce the humidity left
from the decontamination phase.
•
Temperature display shows chamber temperature during cycle.
•
The Text displays that the Decontamination cycle is in progress
•
A timer tracks the remaining time in Decontamination.
COOL DOWN
3 hours
When the humidity reduction period is complete the heaters are shut off to cool the chamber to the user chosen
temperature set point.
•
Temperature display shows chamber temperature during cycle.
•
The Text displays that the unit is now in the Cooling Down cycle
•
A timer tracks how long the unit has been cooling
CYCLE COMPLETE
Indicates the user should:
•
Refill the water pan with single distilled water no purer than 1 Mega OHM in preparation to return to normal
running.
Note:
If Decon Cycle is run as routine maintenance, it is ok to reuse the HEPA filters for the service life indicated in
Section 10. If contamination is an issue due to the room environment, HEPA filters can be replaced as necessary.
RESUME NORMAL OPERATION
Press the Exit button after fulfilling the on screen instructions.
•
This puts the Incubator into normal run mode
•
CO
2
control resumes
•
Checking the calibration of the temperature sensor is recommended.
•
Follow the instructions in section 9.1 of the manual
Содержание NU-5710
Страница 9: ...OM0249 Page 9 of 64 Rev 4 June 2020 ...
Страница 13: ...OM0249 Page 13 of 64 Rev 4 June 2020 ...
Страница 24: ...OM0249 Page 24 of 64 Rev 4 June 2020 ...
Страница 26: ...OM0249 Page 26 of 64 Rev 4 June 2020 ...
Страница 58: ...OM0249 Page 58 of 64 Rev 4 June 2020 ...
Страница 59: ...OM0249 Page 59 of 64 Rev 4 June 2020 ...
Страница 60: ...OM0249 Page 60 of 64 Rev 4 June 2020 ...
Страница 61: ...OM0249 Page 61 of 64 Rev 4 June 2020 ...
Страница 62: ...OM0249 Page 62 of 64 Rev 4 June 2020 ...
Страница 63: ...OM0249 Page 63 of 64 Rev 4 June 2020 ...
Страница 64: ...OM0249 Page 64 of 64 Rev 4 June 2020 ...