8.8.1.3 Default Set points
Temperature 37.0 C
All Models
CO₂
5.0%
Note 1:
A set point of 0 (zero) will put the system into an Idle
state where it does not control and all alarms are
inhibited. The sensor remains active and reports the
level of gas/RH in the chamber.
All Models
RH
(Humidity)
90.0%
Note 1:
A set point of 0 (zero) will put the system into an Idle
state where it does not control and all alarms are
inhibited. The sensor remains active and reports the
level of gas/RH in the chamber.
NU-5720
and
NU-5741
O₂
21.0%
Note 2:
A set point of 21.0% will put the system into an idle
state where it does not control and all alarms are
inhibited.
Note 3:
O₂ set points below 10% will automatically reduce
the Air Injection Time value to “0”. This stops air
injections that will interfere with achieving these
lower O₂ set points.
NU-5731
and
NU-5741
8.8.1.4 Automatic Adjustment of Post Inject Delay Times when
O₂
control system is active
Post inject delays are automatically adjusted according to the
O₂
set point to facilitate gas and RH recovery
after a door opening. The value displayed in the option menu will not change but is the basis for a formula in
the software that calculates the new delay time depending on the
O₂
set point. As the set point is reduced
the delay times are shortened automatically through the software.
The value of the post inject delay time used is the value that shows in the Options display for that control
system whether it is the default or user selected.
8.8.1.5
When the CO₂ system is active the level of CO₂ inversely affects the O₂ present in the chamber.
The maximum achievable O₂ set point is indicated by the following table.
CO₂ Set
Point
Automatic maximum
O₂ Adjustment
0.1% to 1.0%
20.7%
1.1% to 3.0%
20.0%
3.1% to 6.0%
19.0%
6.1% to 10.0%
18.0%
10.1% to 15.0%
17.0%
15.1% to 20.0%
16.0%
Содержание NU-5710
Страница 9: ...OM0249 Page 9 of 64 Rev 4 June 2020 ...
Страница 13: ...OM0249 Page 13 of 64 Rev 4 June 2020 ...
Страница 24: ...OM0249 Page 24 of 64 Rev 4 June 2020 ...
Страница 26: ...OM0249 Page 26 of 64 Rev 4 June 2020 ...
Страница 58: ...OM0249 Page 58 of 64 Rev 4 June 2020 ...
Страница 59: ...OM0249 Page 59 of 64 Rev 4 June 2020 ...
Страница 60: ...OM0249 Page 60 of 64 Rev 4 June 2020 ...
Страница 61: ...OM0249 Page 61 of 64 Rev 4 June 2020 ...
Страница 62: ...OM0249 Page 62 of 64 Rev 4 June 2020 ...
Страница 63: ...OM0249 Page 63 of 64 Rev 4 June 2020 ...
Страница 64: ...OM0249 Page 64 of 64 Rev 4 June 2020 ...