OM0249
Page 30 of 64
Rev 4 June/2020
The paragraphs on the following page provide a more complete description of the diagnostics listed above.
The Diagnostic screen is shown to the right of each of the control systems Temperature, CO
2
, RH & O
2
and the General
category for your reference. Once entered, the Diagnostics screen will display the calibrated and uncalibrated values of the
given function and all related components.
Temperature Control Diagnostics
1.
Safety Relay
This function shows the current state of the safety relay.
LCD display will show “On” or “Off” corresponding to the
relay condition. On for closed or set or yes.
2.
Chamber Temperature Sensor
This function shows on LCD the current values, calibrated (displayed) and
not calibrated of the temperature sensor controlling the chamber
temperature.
3.
Safety Temperature Sensor
This function shows on LCD the current values, calibrated
and not calibrated of the Safety temperature sensor located in the chamber.
4.
Chamber Heater Triac Control
(0%, 50%, and 100%)
This function allows the jacket heater to be turned on at different percentages 0, 50,100)
5.
Door Heater
This function shows on LCD the current state of the door heater. This function also allows the door heater to be
turned on at different percentages (0, 50, and 100 percent). The LCD display will show phase firing percentage
chosen default is 0%.
6.
Perimeter Heater
This function shows the current state of the perimeter heater. This function also allows the perimeter heater to be
turned on at different percentages (0, 50, and 100 percent). The LCD display will indicate percentage of phase
firing chosen default is 0%.
CO
2
Control Diagnostics
1.
CO
2
Sensor
This function shows the current CO
2
sensor reading both the
“Un-Calibrated sensor signal and the Calibrated value shown in the
display screen.
2.
CO
2
Inject Valve /
on/
off
This function shows on LCD the current state of the CO
2
inject valve. LCD
will show “ON” or “OFF” corresponding to the valve condition. On for
open or set or yes.
3.
Sensor Bay Heater Triac Condition /
on/
off
This function shows on LCD the current state of the sensor bay Heater
triac. LCD will show 0, 50, and 100 percent corresponding to the relay
position. The triac can be toggled to each percent of phase firing manually.
4.
CO
2
Sensor Bay Temperature Sensor
This function shows on LCD the sensor bay temperature in degrees C.
The Sensor bay is heated to temperature set point +3.5 degrees
5.
Air inject Solenoid /
on/
off
This function shows on LCD the current state of the Air Inject Solenoid.
LCD will show “ON” or “OFF” corresponding to the solenoid position.
Default is off-solenoid activates when the button is touched
6.
Air Pump /
on/
off
Содержание NU-5710
Страница 9: ...OM0249 Page 9 of 64 Rev 4 June 2020 ...
Страница 13: ...OM0249 Page 13 of 64 Rev 4 June 2020 ...
Страница 24: ...OM0249 Page 24 of 64 Rev 4 June 2020 ...
Страница 26: ...OM0249 Page 26 of 64 Rev 4 June 2020 ...
Страница 58: ...OM0249 Page 58 of 64 Rev 4 June 2020 ...
Страница 59: ...OM0249 Page 59 of 64 Rev 4 June 2020 ...
Страница 60: ...OM0249 Page 60 of 64 Rev 4 June 2020 ...
Страница 61: ...OM0249 Page 61 of 64 Rev 4 June 2020 ...
Страница 62: ...OM0249 Page 62 of 64 Rev 4 June 2020 ...
Страница 63: ...OM0249 Page 63 of 64 Rev 4 June 2020 ...
Страница 64: ...OM0249 Page 64 of 64 Rev 4 June 2020 ...