Run Mode Operator Interactions
8.7.1
THE FIRST TIME THE INCUBATOR IS TURNED ON
•
The NuTouch boot up screen appears then shows a screen instructing you to “Read the manual and follow all
instructions to properly install the incubator” Both screens will automatically advance after 5 minutes. To
manually advance:
o
Press the exit button then a thank you screen will appear.
o
Press the “Continue” button then the unit will go to the “MAIN SCREEN”.
These screens will not show again unless you “MASTER RESET” the incubator control board.
•
Setting the Clock
8.7.1.1
Clock display organization:
YYYY-MM-DD Hr: Min and is displayed in green text in the lower left of the display screen.
8.7.1.2
Press the green text and the Time and date adjustment screen appears.
After a power interruption the date and time will flash as follows:
•
Green text indicates there was a power interruption, but the real time clock remained powered.
Press the flashing text to bring up the “set the clock” screen. If no change is needed press cancel
to go back to the main screen. Otherwise make the required changes then, press the save
button to save the changes and return to the main screen.
•
Yellow text indicates the real time clock lost the backup power and the date and time need to be
reset. Press the flashing text to bring up the date/time setting screen
8.7.1.3
Use the up/down arrows to increase or decrease the highlighted value.
Use the left/right arrows to move between date and time values that require changing.
8.7.1.4
Press “Save” to save the changes you just made or
Press “Cancel” to abort the changes and return to the main screen.
The mode display, on the ‘Main’ screen, is used to identify the mode of chamber operation: either standby or run.
In general, there is no need for operator interaction in "run" mode. However, operator interaction is required when
the incubator is first turned on then to perform calibration functions and address abnormal conditions status. If an
abnormal condition has, or does, exist for a particular parameter, an alarm becomes active and the ‘Alarm Status’
button will appear. This could be a catastrophic alarm condition, which could harm the tissue culture cells. It does
indicate an operational abnormality and should be checked. To acknowledge the abnormality, simply press the
“Alarm Status’ button to view the alarm, to dismiss it press the corresponding alarm’s button. The audio and visual
alarms will then extinguish. If the abnormality still exists, the return after waiting the specified alarm time.
Let the Incubator run normally, and if the alarm doesn't come back on, then everything is normal. If the alarm does
come back on, use the troubleshooting guide to correct the abnormality (see Section 11.0).
8.8
Standby Mode Operator Interactions
The NuAire incubator automatically enters the standby mode when necessary.
There is no action required by the user to enter this mode.
All functions that put the incubator in standby mode automatically start a 10 minute timer.
If the function is not interfaced through the screen on the NuTouch display, the incubator then returns to the MAIN SCREEN
and switches to RUN mode.
8.8.1
Chamber Temperature, C
O₂%, Relative Humidity %, & O₂% Set
- points
8.8.1.1 Accessing the Environmental Set point screen used to manage set
points is done one of two ways:
•
Press any control system value displayed in the main screen.
•
Press the NuAire logo then press the Environmental Set
Point text in the Systems Settings screen
8.8.1.2 Pressing the blue numerical value next to the control system that
you intend to change will bring up a number key pad. Punch in the
new value and hit the
“ent” key to leave the key pad and return to the Set-point screen.
You will be asked if you want to save the new set point(s) when you
press Exit to leave the set point screen and return to the Main screen.
Press “Save” to retain the changes.
Содержание NU-5710
Страница 9: ...OM0249 Page 9 of 64 Rev 4 June 2020 ...
Страница 13: ...OM0249 Page 13 of 64 Rev 4 June 2020 ...
Страница 24: ...OM0249 Page 24 of 64 Rev 4 June 2020 ...
Страница 26: ...OM0249 Page 26 of 64 Rev 4 June 2020 ...
Страница 58: ...OM0249 Page 58 of 64 Rev 4 June 2020 ...
Страница 59: ...OM0249 Page 59 of 64 Rev 4 June 2020 ...
Страница 60: ...OM0249 Page 60 of 64 Rev 4 June 2020 ...
Страница 61: ...OM0249 Page 61 of 64 Rev 4 June 2020 ...
Страница 62: ...OM0249 Page 62 of 64 Rev 4 June 2020 ...
Страница 63: ...OM0249 Page 63 of 64 Rev 4 June 2020 ...
Страница 64: ...OM0249 Page 64 of 64 Rev 4 June 2020 ...