21
Configure Device
|
Novak Electronics, Inc.
www.teamnovak.com
C
C
O
O
N
N
F
F
I
I
G
G
U
U
R
R
E
E
D
D
E
E
V
V
I
I
C
C
E
E
The Novak Data Acquisition software allows you to configure and calibrate the Novak Sentry Data Logger
device. Configure the device settings via the
Device Settings
dialog box, and calibrate the device via the
Customize ESC Input Signal
dialog box.
The Novak Sentry Data Logger must be connected via USB to the computer. Refer to the
USB Connection
section for more details.
You can do one of the following to enter the configuration section:
Click on the
icon next to
Configure Device
when the
connection menu display appears.
On the menu bar, select
Tools>Configure Device
.
On the right lower corner of the application display, click on
the arrow next to the
icon. Select
Configure Device
.
On the tool bar, click on the
Configure Device
icon.
DEVICE SETTINGS
Type a new value over the existing setting, or click on the up and down arrow to change the value.
Changing the Maximum Number of Runs or Sample Rate will change the Maximum Record Time per Run. The
software automatically optimizes the Maximum Record Time per Run to the longest time possible. You can
reduce this time but cannot increase it without editing the Maximum Number of Runs or Sample Rate.
Shown are default Values
Maximum number of Runs
– The maximum number of data files allocated on the
device. Each run requires an available data allocation. The fewer the number of
runs selected, the longer those runs can be.
Maximum Record Time per Run
– The maximum time, in seconds, that can be
logged per run, given the number of runs and sample rates selected. The device will
automatically stop after the configured time.
Sample Rate
– This setting sets the rate speed at which the readings are taken and
recorded from the sensors. The faster the data is recorded (lower number), the
more information will be available for data analysis. However, the faster sample rate
means shorter record time. The slower the rate means the less detailed the
information will be but allows for a longer record time.
WARNING!
Any time a configuration change is made to the Novak Sentry Data Logger device, it will clear
device memory. If there is any data on the device that has not been downloaded, it is advisable to download the
data before making any configuration changes
.
Содержание DATA LOGGER
Страница 1: ...Instruction Manual...