Chapter 3: Tuning and Testing Procedures
© 2019 Cutting Edge Optronics
13
Patara II and Iklwa II Service Manual
Aperture Removal
Figure 2-1: Aperture Assembly (AA21)
1.
Remove the 8-32 screws and locking hardware holding the aperture mount to
the laser bench. Care should be taken to clean any debris from the inside of
the laser housing that may be generated from breaking the epoxy bonds.
2.
Place the aperture assembly on a flat surface. If applicable, heat the epoxy on
the 4-40 screws of the aperture mount with a heat gun and remove the epoxy.
Remove any residual epoxy from the mount.