Lower the current to threshold and check the bullseye on the face of the power
detector to verify proper alignment and operating parameters.
Leak Test
Figure 3-7 Nitrogen Leak Test Layout
1.
Ensure the laser has been purged of any coolant.
2.
Attach a nitrogen source to the
COOLANT IN
port on the laser (see
Figure
3-7
).
3.
Open the inlet valve and adjust the regulator to obtain 80 psi on the pressure
gauge.
4.
Close the inlet valve. Wait 5 minutes and repeat step 3 once.
5.
Return after 30 minutes and record pressure on the gauge. Subtract this
pressure from starting pressure of 80 psi to get the pressure drop over 30
minutes.
6.
Pass/Fail requirements: if the pressure drop is greater than 4 psi the leak test
fails, if the pressure drop is less than 4 psi the leak test passes.