Introduction
to the X-ray system
XT V 160 X-ray System
XTM0011-F2
31
X-rays should always be turned off before opening the operator door;
do not rely on the door interlocks to switch X-rays off.
11.
Manipulator Map (Not shown on overview)
The
Manipulator Map
floating window can be displayed from the
imaging control
toolbar.
It shows an X-ray image-based map (on XT V systems) which can be used to
position the manipulator.
The manipulator map may appear next to the image window,
depending on the system settings.
4.4.1
The Inspect-X manipulator and imaging controls
The controls that you need to position your sample and adjust the image so that you can see
appropriate detail are presented on Inspect-X's
Position and Optimise
tab. This tab is
available to you in the
Radiography, CT
and
X.Tract
workflows. The following section
describes how to:
•
Position the sample
(on page 31)
•
Optimise the image
(on page 32)
•
Switch X-rays on and off
(on page 33)
For more information, refer to the
Inspect-X Online Help
or
User
Manuals
.
Position the sample
To inspect a sample, move it into the path of the X-ray beam using the manipulator. Control
the manipulator axes using:
•
The joysticks on the operator console.
•
The
Axes
controls and readouts in the
Position
panel on the
Position and Optimise
tab.
•
Saved manipulator position sets.
•
The manipulator map (on XT V systems).
•
Interactions with the live image window.
When using the joystick controls, the movement speed can be set with the
Joystick speed
controls. The automatic setting scales the movement speed with the movement of the
joystick, that is, the movement speed is proportional to the amount is a joystick is moved.
There are some small differences to the
Axes
controls in Inspect-X dependent on the type of
manipulator within the X-ray system. In particular, there are different numbers of axes that
Содержание XT V 160
Страница 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 22: ......
Страница 26: ......
Страница 32: ......
Страница 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Страница 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Страница 93: ......