Important health and safety information
2
XTM0011-F2
XT V 160 X-ray System
This system is a Class A product and should not be connected to the
domestic electrical supply. In a domestic environment, this product may
cause harmonic interference. If the user wishes to connect to a
domestic supply they may be required to take adequate measures to
prevent interference.
Machines ship configured for a UK style 230 V (50/60 Hz) AC supply to
a single live conductor, a neutral conductor referenced to earth, and a
separate earth (TN-S). Other supply earth arrangements are not
supported.
The user must not configure or modify the electrical supply in any way.
Using domestic electrical fittings or inadequate wiring may lead to
fitting/wire overheating, failure, or electrical fire.
This system has high protective conductor currents
3.5 mA < I
L
< 10 mA.
It must be connected via either permanent fixed wiring or a suitable
industrial plug and socket.
1.2
Intended use
Nikon Metrology X-ray products are intended for non-destructive inspection of industrial and
electronic components using X-rays.
Do not
use the system for any purpose other than its intended use.
1.3
In case of emergency
To make the system safe in case of emergency, push the emergency
stop button and isolate the system from mains power. The emergency
stop button is located on the control console.
1.4
Health and safety information
1. The relevant “Health and Safety at Work” authorities, for your country, may need to be
informed that you are in possession of equipment designed for the generation of X-rays.
This is a statutory requirement in many countries, including the UK.
2. You must comply with all legislation for X-ray generating equipment existing in your
country. In some countries, both national and local government regulations apply. For
example, in the USA the use of X-ray equipment is controlled by both Federal and State
regulations. In the UK, the applicable regulations are the Ionising Radiation Regulations
2017 (IRR17).
3. It is essential that the personnel engaged in the operation and maintenance, and any
repair of this equipment, be competent and experienced in this work, and that they
comply with the relevant statutory requirements and regulations.
Содержание XT V 160
Страница 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 22: ......
Страница 26: ......
Страница 32: ......
Страница 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Страница 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Страница 93: ......