6.2.6
Fuse list
ID
Type
Function
ID
Type
Function
F2
8A gG 10x38
230V AC T1 O/P
F24
2A T 5x20
230V AC DETECTOR
PSU
F3
10A aM 10x38
75V AC T1 O/P
F25
5x20
230V AC SPARE
F4
10A aM 10x38
75V AC T1 0/P
F26
3.15A T
5x20
+24V SR
F5
6A DC 10x38
100V DC SUPPLY F27
6.3A T 5x20 +24V SOURCE
F6
5A T 5x20
230V AC PSU1
SUPPLY
F28
4A T 5x20
+48V SOURCE
F7
5A T 5x20
230V AC PSU2
SUPPLY
F29
5A T 5x20
24V DC
INTENSIFIER
OPTION
F8
5A T 5x20
230V AC
BACKING PUMP
F20
1A T 5x20
24V DC PSU5 LASER
POINTER
F9
3.15A 5x20
230V AC
COOLANT PUMP
F21
1A T 5x20
24V DC IRIS & JSB
SUPPLY
F20
3.15A 5x20
230V AC
MONITOR
F22
2A T 5x20
24V DC COOLING
TRAY FANS
F21
3.15 A T 5x20
230V AC
AQUISITION PC
F23
2A T 5x20
24V DC ELECTRICAL
AREA FANS
F22
5A T 5x20
230 VAC
PSU3 SUPPLY
F24
2A T 5x20
24V DC
MANIPULATOR
AUX.
F23
2A T 5x20
230V AC
CABINET LIGHT
F25
2A T 5x20
24V DC CT ARM
F26
500mA T
5x20
24V DC XODIAC
MODULE
6.2.7
Radiation integrity
An annual radiation survey on the system is required.
A radiation survey must be performed if the system is moved from one
location to another or if any work is undertaken on the system which
breaks the integrity of the cabinet lead shielding.
Содержание XT V 160
Страница 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 22: ......
Страница 26: ......
Страница 32: ......
Страница 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Страница 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Страница 93: ......