Getting
started
38
XTM0011-F2
XT V 160 X-ray System
Home the manipulator
Before the manipulator in the inspection system can be used, it must be homed. This is
necessary whenever Inspect-X is restarted. The homing operation sends each axis to its
zero position and sets the allowable travel for that axis. Once this operation is complete for
all axes, then the manipulator is calibrated and ready for use.
To
avoid damage to either the manipulator or other materials, ensure
that there are no samples, tools, or other objects in the cabinet that
may collide with the manipulator before you perform a homing
operation.
To home the manipulator, follow these steps:
1. Click
Setup
to select the setup controls.
2. In the
Manipulator
panel, click
.
A Warning appears reminding you to remove from the cabinet any objects that are likely
to collide with the manipulator.
3. Click
OK
at the prompt to proceed.
No other operation can be performed using the manipulator until the homing operation
has completed, or has been stopped by pressing the
Stop
button in the bottom right
corner of the screen.
Whilst the homing operation is in progress, a warning displays:
When homing is complete, the following message appears:
The manipulator is now ready for use.
5.4.2
Load the sample
Move to the load position
Optionally, to make it easier to load the sample into the cabinet, you can move the
manipulator to the
load
position.
To move to the load position, follow these steps:
1. Ensure the manipulator has been homed.
Содержание XT V 160
Страница 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 22: ......
Страница 26: ......
Страница 32: ......
Страница 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Страница 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Страница 93: ......