Chapter 7 Assembly of the Devices
148
7.8.6 Mounting
a
Lamphouse for Dia-illumination
The TI2-D-LHLED LED lamphouse for dia illumination or the D-LH/LC precentered lamphouse to the dia
illumination pillar can be attached. (Tool: 2 mm hexagonal screwdriver, supplied with the microscope main
body)
CAUTION
•
Attachment/removal of a lamphouse
Before attaching or removing a lamphouse, make sure that the light source is turned off.
•
Precautions against heat:
When a precentered lamphouse is used, do not touch the lamphouse while the lamp is lit and for
approximately 30 minutes after it has been turned off.
The basic mounting procedure is common to all dia-illumination lamphouses.
1.
Insert the lamphouse into the lamphouse
mount on the top of the illumination pillar.
Insert the lamphouse by matching the grooves
of the lamphouse cylinder with the pillar-side
pins.
2.
Insert a hexagonal screwdriver from the
hole on the top of the illumination pillar, and
firmly affix the lamphouse by tightening
lamphouse fixing screw.
3.
Affix the lamphouse cable by using the
cable clamp on the rear of the illumination
pillar.
The cable clamp is affixed to the holes of the
illumination pillar through the stoppers at both
ends of the clamp. The cable clamp can be
removed by squeezing the stoppers on both
sides to remove them.
Up to four cables can be routed through the
cable clamp.
Cable connection
Nikon recommends connecting all cables
together at the end of the assembly. For details
on the connector positions, see “7.15
Connection of Each Cable.”
Attaching a lamphouse
Affixing a cable
Cable clamp
Insert a hexagonal screwdriver to tighten the
lamphouse fixing screw.