Chapter 7 Assembly of the Devices
137
7.6 Attaching a Nosepiece
Attach the nosepiece.
The basic mounting procedure is common to all manual nosepieces.
Attaching an objective
Before attaching an objective, make sure that the stage has been mounted.
Mounting a protection plate (optional)
The TI2-N-PT nosepiece protection tray (water tray and protection tray) can also be attached between the
nosepiece and the FL turret. The nosepiece protection tray should be mounted by Nikon personnel. For
details, contact your local Nikon representative.
1.
Remove the drip-proof cover from the top of
the nosepiece.
(Tool: 2 mm hexagonal screwdriver, supplied with
the microscope main body)
Remove the cover mount screw using a hex-
agonal screwdriver, and then remove the
drip-proof cover from the nosepiece.
2.
Attach the nosepiece to the microscope
main body in a correct direction.
Place the nosepiece on the mounting surface of
the elevating junction unit, and then press the
nosepiece from the front of the microscope
toward the rear, until it reaches the limit posi-
tion.
3.
Use a hex wrench to tighten the two M5 hex
socket head cap screws that are incorpo-
rated into the nosepiece.
(Tool: 4 mm hex wrench, supplied with the
microscope main body)
4.
Attach the drip-proof cover on the
nosepiece.
Place the drip-proof cover on the nosepiece,
and then affix the cover mount screw removed
in step 1 by tightening it using a hexagonal
screwdriver.
Attaching a nosepiece
Nosepiece
Drip-proof cover
Cover mount screw
Hex socket head cap screws (M5, x
2)
(built into the nosepiece)