MRT10*RDA001A/E
§6 REPLACEMENT
6 - 1
* For the marked (
) procedures in this section, see the adjustment precautions in §2 SAFETY
PRECAUTIONS.
6.1 Battery AS
Replacement part: 34085-1000
1. Remove the covers. (See 5.1.)
2. Remove the MB main board (34085-BA01).
(See 6.17.)
3. Replace the right side of the battery AS.
* Viewed from the patient’s side
1) Remove connector J3 on the motor CPU
board (34085-BA02).
2) Unscrew HH3×3 (n=2) and SB3×8 (n=2).
Replace the right side of the battery AS
(34085-1000).
4. Replace the left side of the battery AS.
* Viewed from the patient’s side
1) Remove connector J3 on the R motor
CPU board (34085-BA04).
2) Unscrew HH3×3 (n=2) and SB3×8 (n=2).
Replace the left side of the battery AS
(34085-1000).
5. Reassemble the parts in reverse order.
**