4 - 21
MRT10*RDA001A/E
4.8 The communication with the auto refractometer or lensmeter
cannot be done.
Use the auto refractometer and lensmeter that accept
the communication with the RT-5100.
Are the auto refractometer and lensmeter that accept
the communication with the RT-5100 used?
No
Yes
Is the communication cable connected properly?
Does the communication cable break?
Is the communication setting of the auto refractometer
and lensmeter correct?
Is the setting of the auto refractometer and lensmeter
printed when communicating with a memory function?
Is the setting that enables the communication with the
auto refractometer and lensmeter selected?
Replace the RB main board (34085-BA21). (See 6.22.)
Is the symptom relieved?
Completion
Connect the communication cable properly.
(Refer to the operator’s manual.)
Replace the communication cable.
Be sure to print the setting of the auto refractometer
and lensmeter prior to the communication when
communicating with a memory function.
Select the setting that enables the communication with
the auto refractometer or lensmeter.
Auto refractometer or lensmeter malfunctions.
Repair the auto refractometer or lensmeter.
Yes
Yes
Yes
Yes
Yes
No
No
Yes
No
No
No
S e l e c t t h e c o m m u n i c a t i o n s e t t i n g o f t h e a u t o
refractometer and lensmeter correctly.
(Refer to the operator’s manual.)
No
Содержание RT-5100
Страница 6: ...1 2 MRT10 RDA001A E ...
Страница 8: ...2 2 MRT10 RDA001A E ...
Страница 10: ...3 2 MRT10 RDA001A E ...
Страница 70: ...MRT10 RDA001A E 7 8 ...
Страница 71: ...MRT10 RDA001A E 8 REFERENCES 8 1 8 1 Wiring diagram ...
Страница 72: ...MRT10 RDA001A E 8 2 8 2 Connector cables 34085 EA01 34085 EA02 34085 EA04 34085 EA11 ...
Страница 73: ...MRT10 RDA001A E 8 3 34085 EA21 ...
Страница 74: ...MRT10 RDA001A E 8 4 34085 E022 ...