MRT10*RDA001A/E
6 - 8
6.9.2 PD sensors
6.9.2.1 PD sensor of the upper brace R
Replacement part: 34085-BA08
1. Remove the covers. (See 5.1.)
2. Remove connector J8 on the MB main
board (34085-BA01).
3. Unscrew the tight screw P (82053-CE06B)
on the right side (viewed from the patient’s
side) and replace the PI sensor board
(34085-BA08).
4. Reassemble the parts in reverse order.
6.9.2.2 PD sensor of the upper brace L
Replacement part: 34085-BA08
1. Remove the covers. (See 5.1.)
2. Remove connector J10 on the MB main
board (34085-BA01).
3. Unscrew the tight screw P (82053-CE06B)
on the left side (viewed from the patient’s
side) and replace the PI sensor board
(34085-BA08).
4. Reassemble the parts in reverse order.
$#
%'$
%'$
$#