4 - 3
MRT10*RDA001A/E
4.2 Initialization is not completed properly.
4.2.1 The battery AS (34085-1000) does not move to
the right and left.
Does the battery AS (34085-1000) move to the right
and left?
Does the R disk AS (34085-1100) turn?
Does the L disk AS (34085-1200) turn?
4.2.2 The R disk AS (34085-1100) does not turn.
4.2.3 The L disk AS (34085-1200) does not turn.
Yes
Yes
No
Yes
No
No
4.2.4 The PD of the MB and convergence is not
initialized properly.
Is the PD of the MB and the convergence initialized
properly?
Is the right disk of the MB initialized properly?
Is the left disk of the MB initialized properly?
Completion
4.2.5 The right disk of the MB is not initialized
properly.
4.2.6 The left disk of the MB is not initialized properly.
Yes
Yes
No
Yes
No
No
Check the wiring of cable MB and replace it.
Is the voltage between A1-A4 pins of connector J1 on
the MB main board (34085-BA01) DC+5V, and the
voltage between A12-A8 pins DC+16.5V?
No
Yes