4 - 17
MRT10*RDA001A/E
4.4 The instrument cannot be operated with the control box.
4.4.1 The instrument cannot be operated with the switches.
Replace the KEY I/F board (34085-BA12). (See 6.20.)
Is the symptom relieved?
Replace the CB main board (34085-BA11). (See 6.19.)
Yes
Completion
No
4.4.1 The instrument cannot be operated with the
switches.
Can the instrument be operated with the switches?
Can the instrument be operated by turning the dial?
Can the instrument be operated with the touch panel?
Yes
No
Yes
4.4.2 The instrument cannot be operated by turning the
dial.
No
Completion
Yes
4.4.3 The instrument cannot be operated with the touch
panel.
No
Содержание RT-5100