4 - 18
MRT10*RDA001A/E
4.4.2 The instrument cannot be operated by turning the dial.
4.4.3 The instrument cannot be operated with the touch-screen panel.
Perform the position adjustment of the touch-screen
panel. (See 7.13.)
Is the symptom relieved?
Yes
Completion
Replace the color LCD module (42203-E016).
(See 6.25.)
No
Is the symptom relieved?
Yes
Completion
No
Replace the CB main board (34085-BA11). (See 6.19.)
Does cable LCD2 (34085-E304) break?
Yes
Replace cable LCD2 (34085-E304).
No
Replace the CB main board (34085-BA11). (See 6.19.)
Replace the rotary encoder (34085-E202). (See 6.22)
Is the symptom relieved?
Yes
Completion
Replace the KEY I/F board (34085-BA12). (See 6.20.)
No
Is the symptom relieved?
Completion
No
Yes