4.3.1 “Error 01: Communication Error CB-MB” appears.
Replace the RB I/F board (34085-BA22). (See 6.27.)
Is the voltage of the following connector on the RB I/F
board (834085-BA22) as described below?
A1-A4 pins of connector J5:
DC+5V
A12-A8 pins of connector J5: DC+15V
Does cable MB break?
Does cable CB break?
Replace the RB I/F board (834085-BA22). (See 6.27.)
Is the symptom relieved?
Replace the CB main board (34085-BA11).
(See 6.19.)
Replace cable MB.
Completion
No
No
No
No
Yes
Yes
Yes
Replace cable CB.
Yes
Does the MB work properly?
4.2 Initialization is not completed properly.