EDH0162En1040 – 06/99
1.14
MM4005
Introduction
1.3.4.3
Power Inhibition Connector
This 9-pin D-Sub connector provides remote motor power interlock capa-
bility. One or more external switches can be wired to remotely inhibit the
motor power in a way similar to the MOTOR
button on the front
panel. The controller is shipped with a mating 9-pin connector installed
that provides the necessary wiring to enable proper operation without an
external switch.
1.3.4.4
Auxiliary Connector
This 25-pin D-Sub connector has two active lines. One is for motor power
status indication and the other for frequency generator output. The fre-
quency generator is controlled by the motion program and has a frequency
range of 0.01 to 5000 Hz.
1.3.4.5
Remote Control Connector
This 15-pin D-Sub connector provides two functions. The first is similar to
the Power Inhibition connector. The two active pins must be short-circuited
for the motor power to be enabled.
The connector’s second function is to provide inputs for the two analog
ports. These ports are two independent 8 bit analog-to-digital converters.
Programming commands allow the user to read and manipulate the informa-
tion provided by these ports.
The controller is shipped with a mating 15-pin connector installed that pro-
vides the necessary wiring to enable the activation of motor power.
1.3.4.6
RS-232-C Connector
This 9-pin D-Sub connector provides an RS-232-C interface to a host com-
puter or terminal. The port has a three-line configuration using a software
(XON/XOFF) handshake. The pinout enables the use of an off-the-shelf, pin-
to-pin cable. The port provides internally the necessary jumpers to bypass
the hardware handshake, if needed.
1.3.4.7
IEEE-488 Connector
This is a standard 24-pin IEEE-488 connector.
1.3.4.8
Power Switch/Entry Module
The power entry module include a standard IEC 320 inlet combined with a
line filter, fuse box and main power switch. The main power switch turns
power on and off to the entire unit, including the stand-by circuit.
NOTE
The MM4005 senses the line voltage and automatically switches between
110 V and 220 V operation. The acceptable voltage ranges are 95 to 32 V
or 195 to 263 V at 48 to 63 Hz.
While familiarizing yourself with the rear panel and its components, leave
the main power switch in the OFF position. Always make certain the power
switch is in the OFF position before plugging in the power cord.
1.3.4.9
Ground Post
The ground post provides an additional chassis ground connection when
needed. The MM4005 controller chassis can be externally grounded, in
addition to or instead of the grounding supplied through the AC cord.
OFF
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание MM4005
Страница 360: ...Section 7 Servo Tuning Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 368: ...Section 8 Appendices Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 405: ...EDH0162En1040 06 99 8 36 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 406: ...Section 9 Index Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...