3.67
EDH0162En1040 – 06/99
MM4005
LP — List program
Usage
IMM
PGM
MIP
Syntax
xxLP
Parameters
Description
xx
[int]
— Program number.
Range
xx
—
1
to
127.
Units
xx
— None.
Defaults
xx
Missing: Error F.
Out of range: Error F.
Floating point: Error A.
Description
This command reads a specified program from RAM and sends it to the
selected communication port (RS232C or IEEE488). During the transmission
no other command should be sent to the controller.
To read a program from the non-volatile memory, first use the MP com-
mand to download its content in RAM.
Returns
Program listing.
Errors
A
—
Unknown message code.
F
—
Program number incorrect.
G
—
Program does not exist.
I
—
Unauthorized command in programming mode.
Rel. Commands
MP
—
Download EEPROM to RAM.
Example
MP |
Copy programs from EEPROM to RAM.
3LP
|
List program number 3.
1PA0
|
Program listing.
… |
… |
… |
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com