-15-
7. Operation
7-1. Operation Procedure
Operation Flow
The corresponding numbered steps are given for items 1
–12 on the following pages.
Amber TROUBLE LED
flashes.
Change the set values.
6. Incorrect sensor unit is
inserted.
Need to change
No need to change
12. Initial startup is completed.
11. Perform zero adjustment.
11. Is this for
toxic/combustible gas
detection?
8. Check the set
values.
No
Yes
6. Self-test
1
Amber TROUBLE LED is off.
Check the tubing for
looseness.
9.10. Check the
sampling gas for flow
rate and airtightness.
Not normal
Normal
11. Is this for low
oxygen detection?
11. Perform 21vol% adjustment.
No
Yes
2. Turn on the base unit
Base unit startup is
completed.
Initial startup
3. Check the analog
output for base unit
1
Low oxygen detection
Check the power
source.
Check the wiring to the
terminal blocks.
No
Yes
Install the sensor unit.
Install the main unit.
5. Turn on the main unit.
4. Is the sensor unit
installed?
Do you detect toxic
or combustible gas?
3. Adjust the analog
output for base unit
No
Yes
Is the main unit off?
Yes
Turn off the main unit.
No
1. Is power voltage
24 VDC±10%?