-41-
Problem
Probable cause
Steps
Reference section
mode.
No alarm contact
output
MNT switch is set to
MNT1 or MNT2
Set the MNT switch to
the center position for
normal operation
mode.
7-5. Maintenance
Mode
Incorrect wiring
Check and rewire.
Securely connect the
wires to the terminals.
6-3. Wiring
Procedure
Analog output is fixed
at 4 mA or 17.4 mA
MNT switch is set to
MNT2
Set the MNT switch to
the center position for
normal operation
mode.
7-5. Maintenance
Mode
Cannot change the
test mode value
(fixed at 0)
Test mode value is
set within the zero
suppression range
(1) Set the test mode
value to the one
greater than the zero
suppression set value,
or
(2) Change the zero
suppression set value.
(1) 7-4.Test Mode
(2) PS-7 gas
detector’s operation
manual for
administrators
Flow-rate icon slowly
spins
Low flow rate due to
clogged filter
element
Replace the filter
element.
9-1. Filter Element
Replacement
Low flow rate due to
pump failure
Replace the sampling
unit.
9-3. Sampling Unit
Replacement
Low flow rate due to
clogged tubing
Remove clog from
tubing.