User Manual-SKY2000-M4 

Thanks for purchasing our product. Please read the user manual carefully before using. 

Keys Operation: 

ON/OFF

: long press for 3s to turn on/off the device, short press to confirm or save data. 

Menu

: in detection mode, press this button to enter into parameter setting mode, and in parameter 

setting mode, press this button to move the cursor. 

Plus

: in parameter setting mode, short press it to select items upward or add value. Long press for 3 

seconds to turn on/off alarm function quickly. 

Minus

: in parameter setting mode, short press it to select items downward or decrease value. Long 

press for 3 seconds to turn on/off air pump quickly. 

Exit

: in parameter setting mode, short press it to go back to previous menu. Long press to shift the 

gas concentration unit quickly. 

(Note: The long press function only work in detection interface. When you change or reset any data, please 

remember to click 

to confirm and save the settings. Remember to turn on the pump when start the 

detection) 

Power On 

Press the “ON/OFF” for three seconds, the detector will be turned on and the pump will start to work. The 
screen  displays  the  following  interface  successively:  Brand  and  Logo,  main  parameter  interface,  initialing 
countdown. The countdown would be 60 seconds, which is to make sure that the sensors are fully activated. 
After 60 seconds, the device enter normal detection interface.   

Zero Calibration 

Zero calibration must be done in clean air. Zero calibration is needed when there is zero shift happens. 

Put the detector in clean air for 3-5 minutes, under the detection interface, press the “PLUS” and “MINUS” 
buttons at the same time for three seconds, the data of all channels will reset to zero automatically.   

Note: Zero calibration of oxygen detector, nitrogen detector and carbon dioxide detector (target gases are 
component of the air), the data will be defaults to 20.9%VOL (O2), 79.1%VOL (N2), 400ppm (CO2). Do not do 
the zero calibration in not clean air or when there is no zero shift happens. 