![MTS Sensors Level Plus SoCLEAN Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/mts-sensors/level-plus-soclean/level-plus-soclean_operation-manual_1852492008.webp)
8
Level Plus
®
SoClean
®
Operation Manual
Internal electronics
All transmitters come with two electronic components of a sensing
element and a board set. Rigid sensing elements are standard on
SoClean
®
. Flexible sensing elements are available upon request. The
board set consists of up to three electronic boards and a display. The
display is capable of displaying the product level, interface level, and
temperature. Designed into the display are five buttons for local setup of
level transmitter using the stylus.
A temperature sensing function is optional with the SoClean
®
transmitter. The temperature sensing device is a Digital Thermometer
mounted inside the transmitter’s outer pipe assembly. The SoClean
®
can
be ordered with 1, 5, 12, or 16 temperature points.
Accessories
MTS also offers a series of displays, housings, converters, and other
accessories, please refer to the ‘Accessories Catalog’, MTS part number
551103.
5.2 Accuracy
For magnetostrictive transmitters inherent accuracy is measured
in terms of non-linearity. Non-linearity is a measurement of any
imperfections in the waveguide that are reflected in the linearity of the
transmitter’s output. MTS tolerances reflect a maximum non-linearity of
± 1mm. MTS is able to achieve such strict tolerances by manufacturing
all of its own waveguide from a proprietary alloy and testing 100% of all
transmitters before shipping.
5.3 Warranty
All Level Plus
®
transmitters come with a two year limited warranty from
the factory shipment date. A Return Materials Authorization (RMA)
number is required and must accompany any transmitter returns. Any
unit that was used in a process must be properly cleaned in accordance
with OSHA standards, before it is returned to the factory. A Material
Safety Data Sheet (MSDS) must also accompany the transmitter that was
used in any process.
5.4 Storage
If storage is required prior to installation, store indoors in a dry environ-
ment at ambient temperature range not to exceed -40°C (-40°F) to 71°C
(160°F).
Important
:
Contact Technical Support or Customer Service for assistance if you
suspect that the transmitter is not working correctly. Technical support
can assist you with troubleshooting, part replacement, and Returned
Material Authorization (RMA) information if required.
Floats
SoClean
®
transmitters offer numerous floats for different applications
such as stainless steel, 3-A sanitary, hastelloy, Teflon, and Nitrophyl for
both product level and interface level. To be able to accurately detect
the interface level there needs to be a difference of at least 0.05 in
specific gravities between the product and interface liquids. For detailed
information about floats, refer to the ‘Accessories Catalog’, MTS part
number 551103.
For assistance with selecting a specific float for your application, please
contact Technical Support with the following information:
»
Specific gravity of liquid(s) being measured
»
Process temperature
»
Process opening size
»
Vessel pressure
For ATEX and IECEx approvals, SoClean
®
transmitters should be used
with a float having an offset weight and made of stainless steel or Has-
telloy C. This allows the float to stay in contact with the pipe to prevent
the buildup of an electrostatic charge. For detailed information about
floats, refer to the ‘Accessories Catalog’, MTS part number 551103.
Non-metalic floats with a projected surface area of less than 5,000 mm²
should only be used in Zone 0, Gas group IIA such as float part numbers
201643-2, 201649-2, 201650-2, 201109, 251115 and 251116. All
other non-metallic floats offered by MTS such as, 251939, 251119 and
251120 should not be used in a hazardous area application.
Outer pipe configurations
The outer pipe is constructed of a variety of configurations. The
SoClean
®
is available in a sanitary pipe. For other pipe options please
consult other MTS options such as RefineMe
®
, Tank Slayer
®
, and/or
USTDII.
1
1
2
2
3
3
4
4
A
A
B
B
C
C
D
D
REV
ECO #
DESCRIPTION
TITLE
SIZE DWG NO.
REV
PRODUCT LINE
SCALE:
BY
DATE
SHEET
OF
UNLESS OTHERWISE SPECIFIED
TOLERANCES
0 PLACES
MM
1 PLACE
MM
ANGLES
0.5
0.1
- THREAD DEPTHS ARE TO MIN. FULL THDS
- DRILL DEPTHS ARE TO FULL DIA.
- REMOVE BURRS AND SHARP EDGES
- DO NOT SCALE PRINT
- MACHINED SURFACE FINISH 3.2uM [125 uIn]
PROPRIETARY DATA
THE INFORMATION AND DESIGN(S) DISCLOSED
HEREIN ARE THE PROPERTY OF MTS SYSTEMS
CORPORATION AND MAY NOT BE USED,
REPRODUCED OR DISCLOSED IN ANY FORM
EXCEPT AS GRANTED IN WRITING BY MTS
SYSTEMS CORPORATION. THIS RESTRICTION
EXCLUDES INFORMATION THAT IS IN THE
PUBLIC DOMAIN OR WAS LEGITIMATELY IN THE
PRIOR POSSESSION OF THE RECIPIENT.
m
MTS SYSTEMS CORPORATION
SENSORS DIVISION
3001 SHELDON DRIVE, CARY, NC 27513
METRIC
FIRST
ANGLE
C
Marketing
1
1
DRAWN BY:
***NO REVISIONS SHALL BE MADE WITHOUT THE NOTIFICATION TO APPROVAL AGENCY(S).***
Order length
Order Length
3/4 in NPT
Adjustable fitting
Pipe diameter
16mm (0.625 in)
74 mm (2.9 in) min
Inactive zone
Tip of transmitter
Tank Flange
(Provided by customer)
127 [5]
minimum
Tank Flange
1 in NPT or BSPP Adjustable
compression fitting, male
127mm [5 in ]
Order Length
Pipe diameter (0.625 in. dia)
constructed of 316l stainless steel
Sanitary process connection
(sanitary cap)
304.8 mm
(12 in.)
Fig. 5: Outer pipe configuration (TB version shown)