5.6 Specifications (Technical data)
Level Output
Measured variable
Product level and interface level
Output signal /Protocol
Modbus RTU, DDA, Analog (4-20 mA), HART
®
Order length
Sanitary Pipe:
305 mm (12 in.) to 7620 mm (300 in.)
Ƥ
Inherent Accuracy
±1 mm (0.039 in.)
Repeatability
0.001% F.S. or 0.381 mm (0.015 in.)
*
(any direction)
Temperature Output
Measured variable
Average and multipoint temperatures (Modbus, DDA)
Single point temperature (Analog, HART
®
)
Temperature accuracy (Modbus, DDA)
±0.2 ºC (0.4 ºF) range -40 ºC (-40 ºF) to -20 ºC (-4 ºF),
±0.1 ºC (0.2 ºF) range -20 ºC (-4 ºF) to 70 ºC (158 ºF),
±0.15 ºC (0.3 ºF) range 70 ºC (158 ºF) to 100 ºC (212 ºF),
±0.5 ºC (0.9 ºF) range 100 ºC (212 ºF) to 105 ºC (221 ºF)
Temperature accuracy (Analog, HART
®
)
±0.28 ºC (0.5 ºF) range -40 ºC (-40 ºF) to 105 ºC (221 ºF)
Electronics
Input voltage
10.5 to 28 Vdc
Fail safe
High, Full scale (Modbus, DDA)
Low, 3.5 mA default or High, 22.8 mA (Analog, HART
®
)
Reverse polarity protection
Series diode
EMC
EN 61326-1, EN 61326-2-3, EN 61326-3-2, EN 61000-6-2, EN 61000-6-3, EN 61000-4-2, EN
61000-4-3, EN 61000-4-4, EN 61000-4-5, EN 61000-4-6, EN 61000-4-8, EN 61000-4-11
Environmental
Enclosure rating
NEMA Type 4X, IP65
Humidity
0 to 100% relative humidity, non-condensing
Operating temperatures
Electronics:
-40 °C (-40 °F) to 71 °C (160 ºF)
Sensing element:
-40 °C (-40 °F) to 125 °C (257 °F)
◊
Temperature element:
-40 °C (-40 °F) to 105 °C (221 °F)
Vessel pressure
Rigid pipe:
2000 psi (138 bar)
Materials
Wetted parts:
316L stainless steel
†
Non-wetted parts:
316L stainless steel, Epoxy coated aluminum
Field Installation
Housing dimensions
Single cavity:
145 mm (5.7 in.) W x by 127 mm (5 in.) D x 109 mm (4.3 in.) H
Dual cavity:
117 mm (4.6 in.) W x by 127 mm (5 in.) D x 206 mm (8.1 in.) H
Stainless steel single cavity:
178 mm (7.1 in.) W x by 135 mm (5.3 in.) D x 153 mm (6 in.) H
NEMA Type 4X:
87 mm (3.4 in.) W x by 124 mm (4.9 in.) D x 132 mm (5.2 in.) H
Mounting
Rigid pipe
Tri-Clamp, 3/4 in. Adjustable MNPT or BSPP fitting
Wiring
Connections
4-wire shielded cable or twisted pair, Daniel Woodhead 6-pin male connector,
4570 mm (180 in.) integral cable with pigtail
Electrical Connections
Single and Dual Cavity
3/4 in. FNPT conduit opening, M20 for ATEX/IECEx version
NEMA Type 4X
1/2 in. FNPT conduit opening
Display
Measured variables
Product level, interface level and temperature
*
Whichever is greater
∆
Contact factory for longer lengths.
◊
Contact factory for specific temperature ranges.
†
Contact factory for alternative materials.
§
Order length equals the measurement range plus
the inactive zone.