13.2.
Specifications of NDIR technology
ArtNr.
Gas
Measuring Range
Resolution Accuracy
T90
10561
CO
CO
2
CH
4
1000...30.000 ppm
40 Vol.%
1000 ... 10.000 ppm
1 ppm
0,01 Vol.%
1 ppm
±10 ppm / 2%
±0,1 Vol.% / 2%
±10 ppm / 2%
40s
10560
CO
CO
2
C
3
H
8
1000...30.000 ppm
40 Vol.%
1000 ... 10.000 ppm
1 ppm
0,01 Vol.%
1 ppm
±10 ppm / 2%
±0,1 Vol.% / 2%
±10 ppm / 2%
40s
13.3.
Electrochemical Sensor, Sensor for Temperatures and Pressure
Electrochemical Sensor
O
2
Long Life
Measuring Range
0 - 25 Vol%
Resolution
0,01 Vol%
Abs. Accuracy
± 0,2 Vol%
Response Time T90
< 20s
Years expected lifetime (@air)
5 Years
Electrochemical Sensor
CO
H2 compensated
Nom. Measuring Range
0 - 10000 ppm
Overload Range
< 20000 ppm
Resolution
1
Accuracy abs. / reading
± 10 ppm
5 % (0 - 10000 ppm)
10 % (> 20000 ppm)
Response Time T90
< 40 s
Option
CO low
Measuring Range
300 ppm
Resolution
0,1 ppm
Accuracy abs. / reading
2,0 ppm / 5 %
Electrochemical Sensor
CO high
Nom. Measuring Range
0 - 4000 ppm
Overload Range
< 20000 ppm
Resolution
1
Accuracy abs. / reading
± 100 ppm
5 % (0
…
4000 ppm)
10 % (> 4000 ppm)
Response Time T90
< 40 s
Electrochemical Sensor
NO
Nom. Measuring Range
0 - 1000 ppm
Overload Range
< 5000 ppm