User Manual
VARIO
luxx
MRU GmbH, D-74172 Neckarsulm
119 / 129
►
Remove the round filter.
►
Dry the round filter, if necessary.
►
Press OK.
A window appears
►
If necessary, repeat the procedure if another message is displayed.
A message appears, as soon as gas duct is dry again
►
Press CONTINUE MEASUREMENT.
A zero point measurement is started automatically.
15.3.
Switching off the analyser in case error
If the touch screen does not respond, you have two options for switching
off the analyser:
►
Press the
⏻
button during flashing LED for 30 sec. minimum
►
After change to continuous lighting, release the
⏻
button
The instrument will be switching off with reset.
Alternatively, you can switch off the analyser as follows:
►
Press the reset button on the right side of the analyser.
See also the illustration in 3.4 Connectors, Page. 14.