NOTE
Please note that a manually initiated zeroing sets a new
time point grid. This depends on the set interval auto-
zero [h] in the menu Settings.
Example:
You have set a zero-point interval [h] of 4 h. The last
automatic zeroing took place at 2 pm. The next auto-
matic zeroing would be at 6 pm. At 4 pm you carry out
a manually initiated zeroing. The time grid is shifted.
The next automatic zeroing takes place at 8 pm.
You can display the time of the next automatic zeroing.
►
Touch
A window appears.
If the automatic zeroing is switched off "--- "appears in the display.
Use last valid zero point
You can set whether the analyser takes a new zero point or continues to
work with the latest valid zero point.
►
Set in menu Device settings, the Setting " Analyser start with" to
"stored zero point".
See also Chapter 5.4 Device settings, Page 27.
No new zero point is taken after switching on.
The zero-point bar is orange in this case.