User Manual
VARIO
luxx
112 / 129
MRU GmbH, D-74172 Neckarsulm
The heating box is supplied with an external power supply unit.
Precautions for use in the extended temperature range
►
Ensure to switch on the analyser at a temperature of +5° or higher, as
the analyser cannot defrost itself.
►
Ensure that T-air at low temperatures is always is plugged in so that
the heater can work properly.
►
Ensure to close all vents in the transport bag during measurement.
Connecting heating box
►
Insert the connection hose into the condensate outlet in the heating
box (2).
►
Insert the connecting hose into the condensate outlet on the analyser
(4).
1
Heating box
2
Condensate outlet in
Heating box
3
Connection hose
4
Condensate outlet on the
analyser
►
Screw the heating box onto the analyser (1).
The spacer bolts have internal threads.
►
Connect the heating box to the external power supply (3).
►
Plug the external power supply into a power socket.
Ensure that the heating box is supplied with power during the entire
measurement.
►
Ensure to close all vents in the transport bag during measurement.