User Manual
VARIO
luxx
28 / 129
MRU GmbH, D-74172 Neckarsulm
Setting
Range
Description
Combustion analysis
ON/OFF
Setting combustion analysis to ON
allows for several calculation per-
formed
by the instrument, which are fuel
type
Negative gas readings ON/OFF
Negative gas readings caused by
temperature drift of a sensor are
suppressed (shown as zero) or dis-
played
Analyser start with
You can set whether the analyser
starts with
•
a new zero point.
•
a stored zero point.
See also chapter Use last va-
lid zero point, Page 57.
Source zero point (gas)
Setting from which source zero-
point gas is to be sucked in for zero-
point measurement.
See also chapter Select
Source zero point (gas), Page
58.
NOTE:
After switching on, the first zeroing
is always via the fresh air nozzle.
λ
accor. Brettschneider
Calculation of the combustion air ra-
tio (
λ
Lambda) according to
“
Brettschneider
Heated hose
temperature
80
–
180°C
176
–
355°F
Heated probe
temperature
120
–
180°C
250
–
355°F
Ref. temperature
Calculation of the values for a
standard state at reference
temperature
Interval auto-zero [h]
See chapter Automatic zero
point, Page 56.