User Manual
VARIO
luxx
106 / 129
MRU GmbH, D-74172 Neckarsulm
13.4.
Gas sampling and conditioning
Max suction range gas pump
350 hPa
gas flow typ
90 l/h
Single Stage Gas cooler
•
Temperature Peltier cooler (during
grid and battery operation)
5° C
Batt operating hours with gas cooler
(1 Batterypack)
1h45min
Batt operating hours with gas cooler
(NiMH Battery)
30min
Condensate removal from gas cooler
•
Humidity supervision and alarm
•
internal gas flow measurement
•
Gas outlet (Vent port)
•
passiv
3 mm
Optional
active (req. for long VENT
lines or pressurized Vent)
13.5.
Gas cooler strategy
The gas cooler strategy depends on the sensors used.
Operating
condition
Battery
Mains Supply Sensor
Vario
luxx
ECS
max. 20°C
or 5°C below
ambient temp.
and min. 5°C
max.20°C
or 5°C below
ambient temp.
and min. 5°C
ECS
sensor
temperature
Vario
luxx
NDIR + ECS
max.20°C
or 10°C below
ambient temp.
and min. 5°C
5°C
ECS
sensor
temperature
only with bat-
tery operation
Vario
luxx
-
only
paramagnetic
4°C
4°C
only
paramagnetic
Vario
luxx
NDIR + O2 ECS
only
5°C
5°C
Содержание VARIOluxx