15.6.
General Instructions for the heated hose line
For the save operation of the sample line it is important that it is installed
correctly. This includes the following points:
•
The bending radius as given below is maintained,
•
The entire heated hose line is rolled out before put into operation,
•
Both ends of the heated hose line are connected correctly,
•
For thermal and mechanical reasons, it shall be supported by a
cable duct along the line,
•
The heated hose line shall be supported close to the analyser cabi-
net and probe housing, for example by using a separate stilt,
•
There are no extreme temperature variations along the line, as the
line’s temperature is regulated depending on a sensor placed in the
mid of the line.
Unrolling the heated hose line
A wrong and careless use of the heated hose line may cause damage. Es-
pecially on very long heated hose lines the right unpacking and rolling
out is important.
Wrong
Right
Don`t tear the heated sam-
pling line out. ->
Cause sharp kink.
Roll the heated hose line
smooth out.