APPENDIX A
Direct Connection (RS-232)
For a direct connection, set the XMT media to RS-232. Connect the ELI 150 Rx to another Mortara
Instrument electrocardiograph, to an E-Scribe Rx, or to ELI LINK with a direct connect serial cable.
In the configuration setting, select matching baud rates for both units. Use 38400 bps for a direct connection
to E-Scribe Rx.
Modem Transmission
For a modem transmission, set the XMT media to modem. Connect the ELI 150 Rx to a standard telephone
jack with the provided phone line cable. Plug the cable into the telephone jack located on the back of the
electrocardiograph and the other end into a telephone wall jack. Confirm telephone number in the
configuration settings.
Modem Initialization
The modem initialization string is country specific. At the time of production, the modem initialization string
is configured for the country of purchase; however, if the unit is relocated to a different country, the modem
initialization string will need to be modified.
From real-time ECG view, select
F6
(
More
) to display the application menu. From the application menu,
simultaneously press
ALT+SHIFT+M
to access the modem initialization string.
From the Configure Modem screen, select
F2
(
+GCI=
) to populate the prefix “
AT+GCI
” of the modem command.
TIP:
“AT+GCI” will be highlighted – cursor is not present – use keypad to enter country code.
The type of modem installed in your electrocardiograph is displayed at the bottom of the Configure Modem
screen. Use the External Country Code List to enter your country code. Select
F1
(
Send
) to change your
country code. “Sending….” and “Command stored” will be displayed. Select
F6
(
Exit
) to return to the
application menu.
External Modem
Before using the external modem, modem command(s) must be entered and two configuration settings must
be defined. In the configuration, select “external modem” for the XMT media and set the baud rate to 57600.
From the Configure Modem screen, enter and send the initialization command as AT&FEQX4&K.
If a country specific code is necessary, the following list of countries require “%T19,0,XX” to be added to the
modem initialization string. Where “XX” is replaced by the country code number in the following table:
40
Содержание ELI 150 Rx
Страница 12: ...User Safety Information Continued x...
Страница 25: ...ELI 150 Rx Left Side Figure 1 2 ELI 150 Rx Rear Figure 1 3 SECTION 1 3...
Страница 26: ...SECTION 1 ELI 150 Rx Base Figure 1 4 4 T5A 250V...
Страница 34: ...SECTION 2 12...
Страница 44: ...SECTION 3 22...
Страница 58: ...SECTION 4 36...
Страница 74: ...APPENDIX B 52...