APPENDIX A
48
Custom ID Download
Custom ID formats are uniquely defined by your facility’s needs. This customized ECG header information is
designed in an ELI LINK or E-Scribe Rx system and downloaded to the ELI 150 Rx. Up to three custom ID
formats can be downloaded.
From the application menu, select
7
(
Custom ID Download
). “Transmission Status” will remain visible for
approximately 10 seconds followed by “Waiting for Response”, “Connected”, and “Custom ID downloaded”.
A return to the real-time ECG view indicates the custom ID download is complete. The custom ID remains
the new header format for all future ECGs until you select a different ID format in the configuration settings.
You may alter the ID format configuration to short, standard, long, or custom based on your patient
demographic entry needs. The custom ID is only deleted upon downloading a new custom ID or on the rare
occasion of downloading software – it will not be lost due to power loss or switching to a different ID format.
TIP:
Upon custom ID download, the ID format will assume the group name as designed in
ELI LINK or E-Scribe Rx.
NOTE:
The site number must be configured in the electrocardiograph and recognized as an
established, valid site number at the E-Scribe Rx before downloading the custom ID.
TIP:
Confirm the baud rate in the configuration settings before downloading the custom ID
from ELI LINK or E-Scribe Rx .
Содержание ELI 150 Rx
Страница 12: ...User Safety Information Continued x...
Страница 25: ...ELI 150 Rx Left Side Figure 1 2 ELI 150 Rx Rear Figure 1 3 SECTION 1 3...
Страница 26: ...SECTION 1 ELI 150 Rx Base Figure 1 4 4 T5A 250V...
Страница 34: ...SECTION 2 12...
Страница 44: ...SECTION 3 22...
Страница 58: ...SECTION 4 36...
Страница 74: ...APPENDIX B 52...