SECTION 3
Patient Hookup
Correct electrode placement is important for acquiring a successful ECG.
A good minimum-impedance pathway will provide superior noise-free waveforms. Good quality silver-silver
chloride (Ag/AgCl) electrodes should be used.
TIP:
Electrodes should be stored in an air-tight container. Electrodes will dry out if not stored
properly which will cause loss of adhesion and conductivity.
To Attach the Electrodes
❶
Expose the arms and legs of the patient to attach the limb leads.
❷
Place the electrodes on flat, fleshy parts of the arms and legs.
❸
Place the electrodes on the inside of each arm (between the wrist and elbow).
❹
Place the electrodes on the inside of each calf (between the ankle and knee).
➎
Place the electrodes at equal distance from the heart and at the same position on each limb.
➏
For female patients, place the precordial electrodes on top of the breast and not under the
breast tissue.
❼
If a limb site is not available, place the electrodes at an equal distance from the torso, and at an
equal distance on a perfused area of the stump.
❽
Attach the electrodes to the skin. A good test for firm electrode contact is to slightly tug on the
electrode to check adhesion. If the electrode moves freely, it needs to be changed. If the electrode
does not move easily, a good connection has been obtained.
14
Содержание ELI 150 Rx
Страница 12: ...User Safety Information Continued x...
Страница 25: ...ELI 150 Rx Left Side Figure 1 2 ELI 150 Rx Rear Figure 1 3 SECTION 1 3...
Страница 26: ...SECTION 1 ELI 150 Rx Base Figure 1 4 4 T5A 250V...
Страница 34: ...SECTION 2 12...
Страница 44: ...SECTION 3 22...
Страница 58: ...SECTION 4 36...
Страница 74: ...APPENDIX B 52...