28
Configuration Page 2
Flash Size
Indicates the ECG storage capacity. Normal flash size is the standard memory capacity. If the optional
expanded memory has been installed, Expanded will be displayed.
ID Format
Defines the format for the patient demographic information prompts. There are three standard formats: short,
standard, or long. A custom ID format can be downloaded from an ELI LINK or E-Scribe Rx system. Please
see Appendix A to download a custom ID.
The short format includes the patient's last and first name, patient ID number, date of birth (automatically
calculates the age), and gender.
The standard format includes the patient's last name, patient ID number, age, height, weight, gender, race,
medication 1, medication 2, and a location field.
The long format is identical to the standard format except that it includes the patient's first name, room, and
comment fields.
NOTE:
Changing the ID format will affect all configured sites.
Auto-Fill ID
When enabled, the device will automatically populate last name, first name, date of birth, age, and gender in
the ID screen if records with matching patient ID are found in the ECG directory.
AC Filter
The ELI 150 Rx removes 60 Hz or 50 Hz interference. The setting you select depends on the line frequency
in your country. Always use the 60 Hz setting in the U.S. If AC interference is present, check to see that the
proper AC filter is selected.
Paper Speed
The writer speed can be configured at 25 mm/s or 50 mm/s for default ECG printouts. For rhythm printouts
and display, speeds of 5 mm/s or 10 mm/s are also available. See Section 3 to change speeds for display or
rhythm printing. Paper speed is printed at the bottom right corner of the ECG printout.
Filter
The ECG plot-frequency filter (or print filter) can be set to 40 Hz, 150 Hz, or 300 Hz. The plot-frequency filter
does not filter the acquired digital record. A 40 Hz plot-filter setting will reduce the noise (40 Hz and higher
frequencies) on the printed ECG, and a 150 Hz plot-filter setting will reduce the noise (150 Hz and higher
frequencies) on the printout; a 300 Hz plot-filter setting will not filter the printed ECG. The filter setting is
printed at the bottom right corner of the ECG printout.
Height/Weight Units
Defines the units of weight and height to either pounds/inches (lb/in) or kilograms/centimeters (kg/cm).
Содержание ELI 150 Rx
Страница 12: ...User Safety Information Continued x...
Страница 25: ...ELI 150 Rx Left Side Figure 1 2 ELI 150 Rx Rear Figure 1 3 SECTION 1 3...
Страница 26: ...SECTION 1 ELI 150 Rx Base Figure 1 4 4 T5A 250V...
Страница 34: ...SECTION 2 12...
Страница 44: ...SECTION 3 22...
Страница 58: ...SECTION 4 36...
Страница 74: ...APPENDIX B 52...