SECTION 3
20
Acquiring Rhythm Strips
Rhythm strips are printed in the format defined in the configuration (3-channel, 6-channel, or 12-channel).
See Section 4 for instructions to configure rhythm leads.
Begin routine rhythm strips by connecting the patient to the ELI 150 Rx and entering the patient identification
information. After the last data entry field from the ID menu is completed, select
F6
(
Done
) to return to the
real-time ECG view. Select
RHY
to begin rhythm printing.
NOTE:
Rhythm printouts are only possible from the real-time ECG view.
NOTE:
Rhythm acquisitions are only printed and not stored in the ELI 150 Rx .
The rhythm activity screen appears as soon as the writer begins printing the rhythm strip. The waveform display
format is similar to the real-time ECG view; however, new function keys are available during rhythm printing.
In addition to manipulating
SPEED
,
GAIN
, and
FILTER
, the user can toggle different lead groups. When the
default rhythm format is set to either 3-channel or 6-channel, the user can change lead groups during printing
by selecting
F2
(
LEADS
). The change in lead groups is apparent on the printout whereas the waveform display
will remain in the default display of 2.5 seconds of Leads I, II, and V1-V6.
During 3-channel rhythm printing, the available lead groups are:
1. Default (user-selected in configuration)
2. I-II-III
3. aVR, aVL, aVF
4. V1-V2-V3
5. V4-V5-V6
During 6-channel rhythm printing, the available lead groups are:
1. Default (user-selected in configuration)
2. I-II-III-aVR-aVL-aVF
3. V1-V2-V3-V4-V5-V6
During rhythm printing, place the writer in Standby mode by pressing
F6
(
STBY
). To continue rhythm printing
for the same patient without advancing to a new page, select
F6
(
CONT
). To stop the rhythm printing, press
Stop
and the writer will automatically form feed in preparation for a new patient’s rhythm
recording or ECG.
Содержание ELI 150 Rx
Страница 12: ...User Safety Information Continued x...
Страница 25: ...ELI 150 Rx Left Side Figure 1 2 ELI 150 Rx Rear Figure 1 3 SECTION 1 3...
Страница 26: ...SECTION 1 ELI 150 Rx Base Figure 1 4 4 T5A 250V...
Страница 34: ...SECTION 2 12...
Страница 44: ...SECTION 3 22...
Страница 58: ...SECTION 4 36...
Страница 74: ...APPENDIX B 52...