With the pellet on a white paper towel add a few drops of water to the pellet. The water running off the
pellet should be purple and then turn a deep iodine color.
The filter should be considered 80% used when the purple color diminishes from the core of the pellet
and 100% depleted if there is no color present.
use this pellet examination method only in typical laboratory conditions. Highly concentrated
solutions or excessive formalin pouring operations will prematurely deplete the pellets and give a false sense of
filter life.
7.1.3
Health Hazard information & Disposing of Filters
Health Hazard Data - Alumina Permanganate Filter Media
Effects of Exposure – The filter media is non-toxic upon oral, skin, and inhalation exposure and is non-
irritant of the skin. Breathing of dust may cause sneezing. Skin may feel dry after contact. The filter
media is an eye irritant.
Emergency Treatment – Flush eyes with large quantities of water and seek medical attention.
Filter Disposal - Dispose of contaminated filters in accordance with your facility, state and federal procedures for
hazardous materials disposal. See the SDS for the chemicals exposed to the filters for further directions.
7.1.4
Resetting/Leveling the Elevation system
Should your station become unlevel for any reason follow the steps below to reset the elevation system
Press and
hold
both the up/down arrows at the same time.
While holding both buttons, the unit will then move to the lowest position and synchronize and level out.
Continue to hold for another 3 seconds and then release both buttons.
Adjust the unit up and down and confirm the unit has remained leveled. Repeat the reset procedure
multiple times if you suspect it has not fully leveled out.
7.1.5
Resetting the Main GFCI
Should the unit trip the Main GFCI the controls the touch screen and auxiliary power outlets will no longer be powered.
First check that your unit has actually tripped by referencing the image below.
Содержание Maestro MB1000 Series
Страница 51: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 51 9 Documentation 9 1 Sample Rough in Drawing...
Страница 52: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 52 9 2 Electrical diagrams...
Страница 53: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 53...
Страница 54: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 54...
Страница 55: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 55 9 3 Plumbing diagrams...
Страница 56: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 56...