The system will drain down and into the collection carboy. The carboy cap is equipped with a sensor and will
indicate via a status icon when the system is full.
The drain hose is supplied long enough to allow the collection dolly to be pulled out from under the knee space
prior to disconnecting. This will create a small fluid trap if left at factory cut length on collection line.
o
If your preference is to eliminate the fluid trap simply trim the line back short enough that the fluid
drains completely into the carboy. You will have to bend down into the knee space to disconnect now.
Unscrew the carboy cap and remove carboy for disposal. If you prefer to wheel the collection caddy to the
disposal site you will have to disconnect fitting and unscrew both the leak and collection sensors shown above.
6.2.14
Integrated Camera System Options
The Maestro is designed around the PathCam camera system but will also accept other popular pathology camera
systems. The Maestro has two (2) standard camera mounting locations on the underside of the hood canopy.
Содержание Maestro MB1000 Series
Страница 51: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 51 9 Documentation 9 1 Sample Rough in Drawing...
Страница 52: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 52 9 2 Electrical diagrams...
Страница 53: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 53...
Страница 54: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 54...
Страница 55: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 55 9 3 Plumbing diagrams...
Страница 56: ...Installation Service User Manual Maestro series MB1000 MAN rev 1 56...