1.0 

GENERAL OVERVIEW 

The 

ACT-3X

 digital panel meter is an extremely versatile instrument.  The user has complete control of the unit configuration.  

Power may be either 115 - 230 Vac (50/60 Hz), or optionally, 12 Vdc or 24 Vdc.  Input signals are accepted (on Channel 1) from 
optical,  proximity,  magnetic,  infrared  or  laser  sensors,  or  direct  TTL  or  external  AC  sources.    A  second  AUXiliary  input 
(Channel 2) may be used for instrument control and remote resetting. There are several remote communication options – RS232, 
USB or Ethernet and the optional remote software can be used to program the unit or display data locally. The unit is suitable for 
panel mounting or bench top use, with convenient screw terminal connections on the rear panel of the instrument. 

When  the  instrument  is  turned  on,  it  displays  all  8s,  then  ―3  rx.x‖,  where  3  is  the  unit  type  (ACT3)  and  rx.x  is  the  software 
revision level, before entering the normal mode of operation.   

The display will show ―- - - - -‖ when a measurement is over range. 

2.0 

INSTALLATION  

The instrument is intended to operate in the following environment:  

Indoor Use Only

 

Installation Category II

  

per IEC 664    

Pollution Degree Level II 

per IEC61010-1 

Measurement Category I 

per IEC61010-1 

Altitude

  up to 2,000 m

 

Temperature 

-10 °C to +50 °C operating per IEC61010-1 

Humidity

 

Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% 
relative humidity at 40°C 

AC Mains Supply

 

100 - 240 Vac ~  +10% 50/60 Hz 8 VA 

DC Supply (Option) 

12 or 24 VDC+10% 6 watts (DC Option) 

 

NOTE:

 

The  instrument  is  designed  to  be  panel  mounted  and  as  such  should  be  considered  as  fixed 
equipment  or  permanently  connected.  If  permanently  connected,  disconnection  from  the  supply 
must  be  possible  via  a  customer  supplied  switch  or  circuit  breaker  rated  at  120V  or  240V 
(dependent  on  local  voltage  supply)  5A  minimum  when  connected  to  an  AC  supply  or  30V  1A 
minimum when connected to a DC supply. This disconnection device must disconnect all current-
carrying conductors. It must be included in the panel installation and should be clearly marked, in 
close proximity to the Unit and easily accessible to the operator.

   

 

The ACT3X is a 1/8 DIN enclosure requiring a 3.58‖ wide by 1.74‖ high (91x44 mm) mounting hole.  Approximately 5‖ (127 
mm) will be required behind the panel.  Refer to Figure 1 below. 

Before  installing,  check  the  power 
supply  requirements  on  the  rear 
panel.    Remove  the  mounting  clips, 
if  fitted, and install the unit into the 
panel  from  the  front.  From  the  rear 
of  the  unit,  install  the  mounting 
clips  into  the  slots  on  each  side  of 
the  unit  and  tighten  the  mounting 
screws against the front panel. 

WARNING:  Do 

not 

over 

tighten the mounting screws. 

 

 

 

Figure 1  Dimensions in Inches [Millimeters] 

 

WARNING: 

Do  not  use  this  instrument  in  any  manner  inconsistent  with  these 
operating  instructions  or  under  any  conditions  that  exceed  the 
environmental specifications stated. 

 

2.1 

Noisy Environments 

These instruments are highly responsive.  They have input ranges up to 999,990 RPM and 250,000 Hz.  They therefore have 
extremely fast input circuitry that may respond to  spurious noise. It is important to provide a clean source of power to the 
units, either AC or DC, and to ensure that the input to the unit is free of spikes or any other high frequency noise.  In noisy 
environments,  it  may  be  necessary  to  supply  power  through  a  filter,  or  alternate  source.    The  inputs  may  also  need  to  be 

Содержание ACT-3X

Страница 1: ...Totalizer Ratemeter User Manual and Reference Guide MONARCH INSTRUMENT 15 Columbia Drive Amherst NH 03031 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www mona...

Страница 2: ...e personal injury or death may occur as a result 7 This instrument may not be safe for use in certain hazardous environments and serious personal injury or death could occur as a result of improper us...

Страница 3: ...4 3 2 3 UP and DOWN BUTTONS 5 3 2 4 RECALL BUTTON 5 4 0 OPERATION 5 4 1 Modes of Operation Channel 1 5 4 1 1 RPM Mode 5 4 1 2 Frequency Mode 5 4 1 3 Scale Mode Ratemeter 5 4 1 4 Single Event Capture S...

Страница 4: ...5 9 11 5 10 11 5 11 12 5 12 12 6 0 SERIAL OUTPUT 12 6 1 Data from the Serial Interface 12 6 2 Serial Commands 12 6 2 1 Run Mode Commands 12 6 2 2 Control Commands 13 7 0 SPECIFICATIONS 14 8 0 OPTIONS...

Страница 5: ...xed equipment or permanently connected If permanently connected disconnection from the supply must be possible via a customer supplied switch or circuit breaker rated at 120V or 240V dependent on loca...

Страница 6: ...ve Hot wire to the terminal marked L and the Neutral Return wire to the terminal marked N NOTE The ground connection is not required as the unit is fully isolated from the mains If the unit is dc powe...

Страница 7: ...nput is also referred to as Channel 2 CH 2 and will accept a TTL input or any signal up to 12Vdc It has a weak pull up internally and can be used with a set of potential free contacts The AUX input is...

Страница 8: ...is another input pulse It is more useful at slow speeds as it appears to be on continuously at higher inputs It gives an indication that a valid input trigger signal is present 3 1 3 MAX and MIN LEDS...

Страница 9: ...1 2 Frequency Mode In the Frequency Mode the unit displays input pulses per second or more commonly Hertz Hz This is the most basic mode of operation The range of measurement in this mode is 0 0833 t...

Страница 10: ...nding on the sense programmed It acts as a gate preventing the input signal from updating the unit 4 2 3 Single Event Capture When programmed as Single Event Capture this input acts as the second inpu...

Страница 11: ...become de energized Thus in the event of a power failure the relays will drop out by default creating the equivalent of an alarm condition 4 5 Analog AO and Current IO Output The ACT 3X has options f...

Страница 12: ...or 14 640 RPM the internal update rate is the period of the input frequency Thus the response of the alarms etc can be seen to be a function of the input Above an input of 244 Hz the alarms respond wi...

Страница 13: ...d scale factor Press RECALL to exit up one level without making a change or press SET to save the change The edit buttons function as follows RESET changes the digit being edited blinking digit UP inc...

Страница 14: ...EGative or POSitive Press the SET button to select the new setting or the RECALL button to exit without saving the settings 5 3 Sets the decimal point on the displayed reading To enter this menu item...

Страница 15: ...a change or press SET to save the change 5 6 2 FULL SCALE Sets the Full Scale value for the Analog output Press the SET button and use the buttons to edit the value as per above Press RECALL to exit u...

Страница 16: ...ent occurs Other information is sent on demand and is covered in the following section The messages are sent as standard ASCII and all messages end with a carriage return CR There is no Line feed sent...

Страница 17: ...mand Category Argument Value eg CH_1 TYPE SCALE 60 0 sets the Scale Mode Scale factor to 60 0 Command Argument eg DECPT NONE sets the Decimal Point to none Command Category Argument Action CH_1 Channe...

Страница 18: ...100 times per second Standard Gate or up to 1000 Fast Gate Dependent on input frequency Input must be measurement rate Scale Mode Scaling Computation Programmable scaling 0 0001 to 99 999 Front panel...

Страница 19: ...preadsheet Software available for download at www monarchinstrument com ROLS W Remote Optical Laser Sensor with 8 foot cable ROS W Remote Optical Sensor with 8 foot cable ROS P 25 Remote Optical Senso...

Страница 20: ...ing the circumference of the shaft you are monitoring This could also be a speed wheel tire etc The circumference x diameter 3 14159 In order to scale we need to know what we want as opposed to what w...

Страница 21: ...il the display shows and then press the SET button The display will show the current scale factor value with the right most digit flashing The UP and DOWN buttons change the flashing digit The RESET b...

Страница 22: ...an error in the placement of the sensors The tape edges won t be exactly 1 inch apart Assuming the edges were really 1 01 inches apart there would be 1 error So at 130 MPH your reading would be 128 7...

Страница 23: ...hat the device is set up for DHCP by default so it will get an IP address from your local server Once the search is complete the window will be populated as shown next The current device setting is sh...

Страница 24: ...ons to set the required value Press RECALL to exit up one level without making a change or press SET to make the change The edit buttons function as follows RESET changes the digit being edited blinki...

Отзывы: