Page 9
Decimal Point
- Set the number of Decimal Points displayed on the unit
Set point 1
- Set up parameters for Alarm 1
Set point 2
- Set up parameters for Alarm 2
Digital Analog Converter 1
- Set up parameters for the Analog Output (Option)
Baud Rate
- Set up the baud rate for the serial port (Option). Only seen if the option is installed.
Ethernet
- Set up the Ethernet port address, subnet mask and gateway. Only seen if the option is installed.
Pulse Output
- Select the polarity of the pulse output
Display Rate
- Select the display update rate
Serial Number
- Display the Serial Number
Passcode
- Display or set the pass code to protect the Menu
When at the desired menu option, press the
SET
button to enter into that specific menu option.
5.1
Sets the mode of operation. This menu item has a number of sub items. Once selected with the
SET
button, the instrument
will display its current mode. To exit back to the main menu without changing the mode, press the
RECALL
button. If you
wish to change the operating mode, you can view the options using the
UP
button. The available options are shown
. When the desired mode is displayed, press the
SET
button. The unit will return to
the main menu.
5.1.1
TYPE -
Press the
SET
button to enter the sub menu selections shown below and use the
UP
button to scroll through the options.
See
5.1.1.1
RPM -
Sets the unit into RPM mode.
5.1.1.2
FREQ -
Sets the unit into Frequency (Hz) mode.
5.1.2
SCALE
-
Sets the unit into Scale mode. User must then specify a scale factor from 0.0001 to
99,999. Press the
SET
button to display the current scale factor value and use the buttons to set the
required scale factor. Press
RECALL
to exit up one level without making a change or press
SET
to save the change. The edit buttons function as follows:
RESET
- changes the digit being edited (blinking digit)
UP
- increases the digit by 1
DOWN
- decreases the digit by 1
RESET
+
- shifts the decimal point
RECALL
- aborts the edit process
SET
- saves data and exits
5.1.2.1
SECAP -
Sets the unit into Single Event Capture mode. User must then specify a scale factor
from 0.0001 to 99,999. Press the
SET
button and use the buttons (as described in SCALE
above) to set the required scale factor value. Press
RECALL
to exit up one level without
making a change or press
SET
to save the change.
5.1.2.2
ROC -
Sets the unit into Rate of Change mode. User must then specify a scale factor from
0.0001 to 99,999. Press the
SET
button and use the buttons (as described in SCALE above) to
set the required scale factor value. Press
RECALL
to exit up one level without making a
change or press
SET
to save the change.
5.1.2.3
TOTAL -
Sets the unit into Totalizer mode. User must then specify a constant value from
0.00001 to 99,999. Press the
SET
button and use the buttons (as described in SCALE above)
to set the required constant value. Press
RECALL
to exit up one level without making a
change or press
SET
to save the change.
5.1.3
INPUT -
Press the
SET
button to change the sense of the input trigger and use the
UP
button to
select NEGative or POSitive. Press the
SET
button to select the new setting or the
RECALL
button to exit without saving the settings.
5.1.4
LOEND -
Visible only in RPM, FREQ, SCALE and SECAP modes. Press the
SET
button to enter
the settings for the
Low End
measurement (how long the unit waits for a measurement before
showing Zero) and use the
UP
button to scroll through the options described below. Press the
SET
button to select the new setting or the
RECALL
button to exit without saving the settings.