background image

Page 6

 

 
 

In this mode: 

 

The display is updated up to once every two seconds with the average Rate of Change. 

 

The throughput of the max/min, analog outputs, and relays is up to twice a second. 

 

The  display  will  show  a  positive  number  when  the  frequency  is  increasing  and  a  negative  number  when  the 
frequency is decreasing. 

Remember that even a small change in RPM over a short time will cause a large Rate of Change (average acceleration) 
to  be  displayed.    The  gate  time  for  each  measurement  is  about  1/3  of  a  second.    For  instance,  if  you  use  a  digital 
function generator to change from 3000 RPM to 3001 RPM, the change will happen all at once.  You will see that it 
happens within 1/3 of a second.  So 1 RPM change in 1/3 of a second is 180 RPM over one minute or 180 PRMPM 
acceleration. 

4.1.6

 

Totalizing Mode 

In the 

Totalizing Mode

, each input pulse causes the display to be incremented by a constant value that is set by the 

user.  This enables the user to scale the input to obtain a read out in any measure required: number of inches, number of 
bottles, number of revolutions, and so on.  The scale factor may be set anywhere from 0.0001 to 99,999. The total can 
be reset using the 

RESET

 and 

RECALL

 buttons pressed together on the front panel. The AUX input (Channel 2) can 

be set to reset the totalizer and turn the unit into a batch counter or to inhibit the input signal (see 

Section 4.2.4

). 

4.2 

Modes of Operation 

– Channel 2 

Channel 2 is the AUXiliary input and is basically used to control the input to Channel 1, or can be used as an external reset 
for the input channel or the Alarms. The user can set the sense of the input, positive or negative. The input has a weak pull-
up resistor internally so it can be used with a potential free contact. All modes are mutually exclusive – only one mode may 
be active at any time. Refer to 

Section 5.0

 for details on changing modes. 

NOTE: 

The instrument is programmed from the factory with the AUXiliary input disabled (OFF). 

4.2.1

 

Off 

When programmed OFF this input has no effect. 

4.2.2

 

Inhibit 

If  programmed  as  an  inhibit  pin,  it  will  disable  the  signal  input  when  asserted  (high  or  low  depending  on  the  sense 
programmed). It acts as a gate preventing the input signal from updating the unit. 

4.2.3

 

Single Event Capture 

When  programmed  as  Single  Event  Capture  (



)  this  input  acts  as  the  second  input  for  the  external  signal  –  see 

Section 4.1.4

 above. 

4.2.4

 

External Reset 

– Input Totalizer 

If programmed as External Reset for Inputs (



), when asserted (high or low depending on the sense programmed), 

it  will  reset  the  display  (count)  to  Zero.  Used  in  the  Totalizer  Mode  (see 

Section  4.1.6

)  to  enable  the  unit  as  a  batch 

counter. 

4.2.5

 

External Reset 

– Alarms 

If  programmed  as  External  Reset  for  Alarms  (



),  when  asserted  (high  or  low  depending  on  the  sense 

programmed), it will reset any Alarms and drop out any Contact closures. Note: The alarms need to be set up correctly 
(see 

Section 4.4

).

 

4.3 

Decimal Point 

The decimal point on the display may be fixed from 1 to 3 places or may be set to none (only whole number displayed). As 
the  display  increases  the  unit  will  automatically  drop  decimals  to  show  whole  number  values.  Note  that  for  values  larger 
than 99999 all decimal points light indicating the current reading is x10. 

4.4 

Alarm Limits 

The 

ACT-3X

 has two independent alarm set points, referred to as 

LIMIT 1

 and 

LIMIT 2 

(

Set 1

 and

 Set 2

 on the menu)

.

  

These limits are fully programmable by the user (unless the write protect option has been set).  The limits may be set as high 
or  low  with  an  option  of  low  limit  lockout,  latching  or  non-latching  at  any  value.    The  limits  are  accurate  to  better  than 
±0.1%  of  the  set  point  value.    Refer  to 

Section  4.7.2

  for  the  limit  response  time.    The  dead  band  (hysteresis)  is  also 

programmable at any value from 0.0001 to 99.999% of the set point value.  The actual output from these alarms is a set of 
form C, dry contacts, accessible via barrier strip screw terminals on the rear panel. These contacts are capable of switching 1 
A at 250 Vac.  When the unit is making measurements, the limits can be viewed by pressing the 

UP 



 button for 

LIMIT 1

 

or the 

DOWN 



 button for 

LIMIT 2.

  The display will return to normal after a few seconds.  The relays may also be set to 

be Failsafe, which inverts the sense of the relay so that they are energized under normal condition. In the event of a power 
failure the contacts will drop out as they would during an alarm condition. Refer to 

Section 5.0

 to set limit set points. 

 

Содержание ACT-3X

Страница 1: ...Totalizer Ratemeter User Manual and Reference Guide MONARCH INSTRUMENT 15 Columbia Drive Amherst NH 03031 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www mona...

Страница 2: ...e personal injury or death may occur as a result 7 This instrument may not be safe for use in certain hazardous environments and serious personal injury or death could occur as a result of improper us...

Страница 3: ...4 3 2 3 UP and DOWN BUTTONS 5 3 2 4 RECALL BUTTON 5 4 0 OPERATION 5 4 1 Modes of Operation Channel 1 5 4 1 1 RPM Mode 5 4 1 2 Frequency Mode 5 4 1 3 Scale Mode Ratemeter 5 4 1 4 Single Event Capture S...

Страница 4: ...5 9 11 5 10 11 5 11 12 5 12 12 6 0 SERIAL OUTPUT 12 6 1 Data from the Serial Interface 12 6 2 Serial Commands 12 6 2 1 Run Mode Commands 12 6 2 2 Control Commands 13 7 0 SPECIFICATIONS 14 8 0 OPTIONS...

Страница 5: ...xed equipment or permanently connected If permanently connected disconnection from the supply must be possible via a customer supplied switch or circuit breaker rated at 120V or 240V dependent on loca...

Страница 6: ...ve Hot wire to the terminal marked L and the Neutral Return wire to the terminal marked N NOTE The ground connection is not required as the unit is fully isolated from the mains If the unit is dc powe...

Страница 7: ...nput is also referred to as Channel 2 CH 2 and will accept a TTL input or any signal up to 12Vdc It has a weak pull up internally and can be used with a set of potential free contacts The AUX input is...

Страница 8: ...is another input pulse It is more useful at slow speeds as it appears to be on continuously at higher inputs It gives an indication that a valid input trigger signal is present 3 1 3 MAX and MIN LEDS...

Страница 9: ...1 2 Frequency Mode In the Frequency Mode the unit displays input pulses per second or more commonly Hertz Hz This is the most basic mode of operation The range of measurement in this mode is 0 0833 t...

Страница 10: ...nding on the sense programmed It acts as a gate preventing the input signal from updating the unit 4 2 3 Single Event Capture When programmed as Single Event Capture this input acts as the second inpu...

Страница 11: ...become de energized Thus in the event of a power failure the relays will drop out by default creating the equivalent of an alarm condition 4 5 Analog AO and Current IO Output The ACT 3X has options f...

Страница 12: ...or 14 640 RPM the internal update rate is the period of the input frequency Thus the response of the alarms etc can be seen to be a function of the input Above an input of 244 Hz the alarms respond wi...

Страница 13: ...d scale factor Press RECALL to exit up one level without making a change or press SET to save the change The edit buttons function as follows RESET changes the digit being edited blinking digit UP inc...

Страница 14: ...EGative or POSitive Press the SET button to select the new setting or the RECALL button to exit without saving the settings 5 3 Sets the decimal point on the displayed reading To enter this menu item...

Страница 15: ...a change or press SET to save the change 5 6 2 FULL SCALE Sets the Full Scale value for the Analog output Press the SET button and use the buttons to edit the value as per above Press RECALL to exit u...

Страница 16: ...ent occurs Other information is sent on demand and is covered in the following section The messages are sent as standard ASCII and all messages end with a carriage return CR There is no Line feed sent...

Страница 17: ...mand Category Argument Value eg CH_1 TYPE SCALE 60 0 sets the Scale Mode Scale factor to 60 0 Command Argument eg DECPT NONE sets the Decimal Point to none Command Category Argument Action CH_1 Channe...

Страница 18: ...100 times per second Standard Gate or up to 1000 Fast Gate Dependent on input frequency Input must be measurement rate Scale Mode Scaling Computation Programmable scaling 0 0001 to 99 999 Front panel...

Страница 19: ...preadsheet Software available for download at www monarchinstrument com ROLS W Remote Optical Laser Sensor with 8 foot cable ROS W Remote Optical Sensor with 8 foot cable ROS P 25 Remote Optical Senso...

Страница 20: ...ing the circumference of the shaft you are monitoring This could also be a speed wheel tire etc The circumference x diameter 3 14159 In order to scale we need to know what we want as opposed to what w...

Страница 21: ...il the display shows and then press the SET button The display will show the current scale factor value with the right most digit flashing The UP and DOWN buttons change the flashing digit The RESET b...

Страница 22: ...an error in the placement of the sensors The tape edges won t be exactly 1 inch apart Assuming the edges were really 1 01 inches apart there would be 1 error So at 130 MPH your reading would be 128 7...

Страница 23: ...hat the device is set up for DHCP by default so it will get an IP address from your local server Once the search is complete the window will be populated as shown next The current device setting is sh...

Страница 24: ...ons to set the required value Press RECALL to exit up one level without making a change or press SET to make the change The edit buttons function as follows RESET changes the digit being edited blinki...

Отзывы: