background image

Page 8

 

 
 

4.7 

Throughput 

Throughput  is  a  measure  of  how  fast  the  instrument  processes  data.    The  rate  at  which  the  instrument  acquires  data  is  a 
function of the ―Gate Time‖ and the input frequency.  The instrument gets a start pulse then it continues to get pulses until 
the Gate Time elapses.  The next pulse ends this measurement and starts the next. At frequencies slower than the Gate Time, 
the update rate is equal to the period of the input frequency.  Eventually, the instrument has to make the decision that the 
input is zero, because theoretically it could wait forever for the next pulse.  This Low-End timeout is programmable.   

4.7.1

 

Display Update Rate 

Although the instrument can update up to 244 times a second, to display the data at this rate would result in a totally 
erratic display.  Therefore, the instrument limits the display update rate to once every ½ second.  Obviously if the input 
pulses  are  spaced  more  than  ½  second  apart,  the  instrument  will  not  have  any  new  data  until  the  next  pulse  comes 
along, and the time to update will be greater than ½ second.   The point at which the update rate becomes longer than 
every ½ second is when the period of the input (time between pulses) is greater than ½ second, which is 2 Hz or 120 
RPM.  Thus, for an input greater than 2 Hz or 120 RPM, the update rate is twice a second.   

For very fast inputs, the unit averages the readings between display updates so that the value displayed is an average of 
the total number of acquisitions since the last update. 

4.7.2

 

Internal Update Rate - GATE 

The rate at which the limits are checked, the analog output is updated, and the minimum and maximum are updated, is 
at the maximum rate at which the instrument acquires data.  This is set by the 

GATE

 menu item. The Gate Time can be 

set to 32.786 mSecs (

STD

) or 4.096 mSecs (

FAST

). See 

Section 5.1.4

 for more details.  

The 

STD

 setting is slower (up to 31 readings per second) but gives more accurate readings especially for the maximum 

and minimum readings.  Below 31 Hz or 1860 RPM, the internal update rate is the period of the input frequency. Thus, 
the response of the alarms, etc can be seen to be a function of the input.  Above an input of 31 Hz, the alarms respond 
within 66 milliseconds.  Below this input they respond within (1 

 input frequency) seconds. 

The 

FAST

 gate time is faster (up to 244 readings per second) but is less accurate (about 0.025% of reading worst case 

at high frequencies). Below 244 Hz or 14,640 RPM, the internal update rate is the period of the input frequency. Thus, 
the response of the alarms, etc can be seen to be a function of the input.  Above an input of 244 Hz, the alarms respond 
within 9 milliseconds. Below this input they respond within (1 

 input frequency) seconds. 

At input frequencies below 31 Hz or 1860 RPM there will be no difference in the two settings. 

The  instrument  has  a  special  feature  to  allow  it  to  quickly  respond  to  rapid deceleration  and  still measure  down  to 5 
RPM with 1 pulse per revolution.  (To  measure to 1 RPM, 5 pulses per revolution are required).  After receiving no 
input pulses for about 67 milliseconds for the Standard gate mode or 37 mSecs for the Fast gate mode, the instrument 
will calculate a reading as though an input pulse had just occurred.  If this new reading is less than the last reading, the 
instrument  uses  it.    Until  an  input  pulse  is  detected  or  the  Low-End  timeout  is  reached,  the  instrument  will  ―force‖ 
another reading every 33 milliseconds. These ―forced‖ readings will update the analog output, limits, and the max/min.  
The last ―forced‖ reading of every ½ second will be displayed every ½ second. The Low-End timeout can be set to 12, 
1 or 0.5 seconds.  Refer to 

Section 5.1.3

 for details. 

4.7.3

 

Low End 

Low End is applicable to 

RPM

FREQ

SCALE

, and 

SECAP

 Modes only. Low End selects how many seconds may 

elapse between input pulses before the unit displays the reading zero.  There is a tradeoff between the lowest reading 
available  and  how  quickly  the  unit  responds  when  the  input  pulses  stop  and  displays  0.    There  are  three  values:  12 
seconds, 1 second, and 1/2 seconds.  In the RPM mode, with one pulse per revolution, these settings correspond to the 
lowest RPM reading of 5, 60 and 120 RPM respectively.  

4.8 

PULSE OUTPUT 

The  Pulse  Repeater  Output provides  a  conditioned  TTL  positive  going  5 V pulse  out  for  each  pulse  in.    The  sense  of  the 
output, high pulse or low pulse, is programmable by the user. See 

Section 5.9

5.0 

USING THE MENU 

To  enter  the 

MENU  Mode

,  press  the 

SET

  button.  The  display  will  show  the  first  top-level  menu  item,  which  is 



  for 

Channel 1 setup. Continuing to press the 

SET

 button will cycle through all options and show the current settings. At any point 

press the 

RECALL

 button to back out (



) the current setting. Pressing 

RECALL

 again will eventually exit from the Menu.  

Once  in  the MENU  Mode,  with 



 

displayed,  press  the 

UP 



 

or 

DOWN 



 buttons to  cycle  through  the  top-level  menu 

options.  

The top-level menu choices are:  



 

Channel 1 

- Set up parameters for Channel 1 – the primary input Channel



 

Channel 2 

- Set up parameters for Channel 2 – the Auxiliary Input Channel

Содержание ACT-3X

Страница 1: ...Totalizer Ratemeter User Manual and Reference Guide MONARCH INSTRUMENT 15 Columbia Drive Amherst NH 03031 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www mona...

Страница 2: ...e personal injury or death may occur as a result 7 This instrument may not be safe for use in certain hazardous environments and serious personal injury or death could occur as a result of improper us...

Страница 3: ...4 3 2 3 UP and DOWN BUTTONS 5 3 2 4 RECALL BUTTON 5 4 0 OPERATION 5 4 1 Modes of Operation Channel 1 5 4 1 1 RPM Mode 5 4 1 2 Frequency Mode 5 4 1 3 Scale Mode Ratemeter 5 4 1 4 Single Event Capture S...

Страница 4: ...5 9 11 5 10 11 5 11 12 5 12 12 6 0 SERIAL OUTPUT 12 6 1 Data from the Serial Interface 12 6 2 Serial Commands 12 6 2 1 Run Mode Commands 12 6 2 2 Control Commands 13 7 0 SPECIFICATIONS 14 8 0 OPTIONS...

Страница 5: ...xed equipment or permanently connected If permanently connected disconnection from the supply must be possible via a customer supplied switch or circuit breaker rated at 120V or 240V dependent on loca...

Страница 6: ...ve Hot wire to the terminal marked L and the Neutral Return wire to the terminal marked N NOTE The ground connection is not required as the unit is fully isolated from the mains If the unit is dc powe...

Страница 7: ...nput is also referred to as Channel 2 CH 2 and will accept a TTL input or any signal up to 12Vdc It has a weak pull up internally and can be used with a set of potential free contacts The AUX input is...

Страница 8: ...is another input pulse It is more useful at slow speeds as it appears to be on continuously at higher inputs It gives an indication that a valid input trigger signal is present 3 1 3 MAX and MIN LEDS...

Страница 9: ...1 2 Frequency Mode In the Frequency Mode the unit displays input pulses per second or more commonly Hertz Hz This is the most basic mode of operation The range of measurement in this mode is 0 0833 t...

Страница 10: ...nding on the sense programmed It acts as a gate preventing the input signal from updating the unit 4 2 3 Single Event Capture When programmed as Single Event Capture this input acts as the second inpu...

Страница 11: ...become de energized Thus in the event of a power failure the relays will drop out by default creating the equivalent of an alarm condition 4 5 Analog AO and Current IO Output The ACT 3X has options f...

Страница 12: ...or 14 640 RPM the internal update rate is the period of the input frequency Thus the response of the alarms etc can be seen to be a function of the input Above an input of 244 Hz the alarms respond wi...

Страница 13: ...d scale factor Press RECALL to exit up one level without making a change or press SET to save the change The edit buttons function as follows RESET changes the digit being edited blinking digit UP inc...

Страница 14: ...EGative or POSitive Press the SET button to select the new setting or the RECALL button to exit without saving the settings 5 3 Sets the decimal point on the displayed reading To enter this menu item...

Страница 15: ...a change or press SET to save the change 5 6 2 FULL SCALE Sets the Full Scale value for the Analog output Press the SET button and use the buttons to edit the value as per above Press RECALL to exit u...

Страница 16: ...ent occurs Other information is sent on demand and is covered in the following section The messages are sent as standard ASCII and all messages end with a carriage return CR There is no Line feed sent...

Страница 17: ...mand Category Argument Value eg CH_1 TYPE SCALE 60 0 sets the Scale Mode Scale factor to 60 0 Command Argument eg DECPT NONE sets the Decimal Point to none Command Category Argument Action CH_1 Channe...

Страница 18: ...100 times per second Standard Gate or up to 1000 Fast Gate Dependent on input frequency Input must be measurement rate Scale Mode Scaling Computation Programmable scaling 0 0001 to 99 999 Front panel...

Страница 19: ...preadsheet Software available for download at www monarchinstrument com ROLS W Remote Optical Laser Sensor with 8 foot cable ROS W Remote Optical Sensor with 8 foot cable ROS P 25 Remote Optical Senso...

Страница 20: ...ing the circumference of the shaft you are monitoring This could also be a speed wheel tire etc The circumference x diameter 3 14159 In order to scale we need to know what we want as opposed to what w...

Страница 21: ...il the display shows and then press the SET button The display will show the current scale factor value with the right most digit flashing The UP and DOWN buttons change the flashing digit The RESET b...

Страница 22: ...an error in the placement of the sensors The tape edges won t be exactly 1 inch apart Assuming the edges were really 1 01 inches apart there would be 1 error So at 130 MPH your reading would be 128 7...

Страница 23: ...hat the device is set up for DHCP by default so it will get an IP address from your local server Once the search is complete the window will be populated as shown next The current device setting is sh...

Страница 24: ...ons to set the required value Press RECALL to exit up one level without making a change or press SET to make the change The edit buttons function as follows RESET changes the digit being edited blinki...

Отзывы: