Page 10

 

 
 

5.1.4.1   



 

12 SEC -

 Sets the Low End to 12 Seconds, equivalent to 5 RPM.

 

5.1.4.2   



 

1 SEC -

 Sets the Low End to 1 Second, equivalent to 60 RPM.

 

5.1.4.3   



 

HALF -

 Sets the Low End to 1/2 Second, equivalent to 120 RPM.

 

5.1.5

 

 



 

GATE -

 Press the 

SET

 button to enter the settings for the 

GATE

 time (internal measurement rate) 

and use the 

UP

 



 button to scroll through the options described below. Press the 

SET

 button to 

select the new setting or the 

RECALL

 button to exit without saving the settings.

   

5.1.5.1   



 

STD -

 Sets the gate to the Standard rate – 32.768 msec.

 

5.1.5.2   



 

FAST -

 Sets the gate to the FAST rate – 4.096 msec.

 

5.2 

 



 

Sets  the  Auxiliary  input  function.    This  menu  item  has  a  number  of  sub  items.    Once  selected  with  the 

SET

  button,  the 

instrument  will  display  its  current  function.    To  exit  back  to  the  main  menu  without  changing  the  function,  press  the 

RECALL

  button.    If  you  wish  to  change  the  function,  you  can  view  the  options  using  the 

UP 



 

button.    The  available 

options are shown below and are described in 

Section 4.2

.  When the desired function is displayed, press the 

SET

 button.  

The unit will return to the last menu level.  

5.2.1

 

 



 

TYPE -

 Press the 

SET

 button to enter the sub menu selections shown below and use the 

UP

 



 

button to scroll through the options.

  

5.2.1.1   



 

OFF -

 Turns OFF the Aux input so that it has no function.

 

5.2.1.2   



 

SECAP -

 Sets the Aux input to be the second input for the SECAP mode. Se

Section 4.2.3

.

 

5.2.1.3   



 

INHIBIT -

 Sets the Aux input to disable the input. Se

Section 4.2.2

.

 

5.2.1.4   



 

RESET INPUT -

 Sets the Aux input to RESET the current Display Total Value. See 

Section 

4.2.4

.

 

5.2.1.5   



 

RESET ALARM -

 Sets the Aux input to be the ALARM reset. Se

Section 4.2.5

.

 

5.2.2

 

 



 

INPUT -

 Press the 

SET

 button to change the sense of the Aux input and use the 

UP

 



 button to 

select  NEGative  or  POSitive.  Press  the 

SET

  button  to  select  the  new  setting  or  the 

RECALL

 

button to exit without saving the settings.

   

5.3 

 



 

Sets the decimal point on the displayed reading. To enter this menu item press the 

SET

 button.  The unit will display the 

current decimal point setting.  To select a different decimal place, press the 

UP 



 

button, which will toggle between 1, 2, 3 

and NONE.  Press the 

SET

 button to change the decimal point setting to the option displayed and exit back to the menu. To 

exit back to the main menu without changing the mode, press the 

RECALL

 button.   

5.4 

 



 

Used to set or show the 

ALARM

 

1

 SETUP parameters.  This menu item has a number of sub items.  Once selected with the 

SET

 button, the instrument will display its current setting.  If you wish to change the alarm setup, you can view the options 

using  the 

UP 



 

button.    The  available  options  are  shown  below  and  are  described  in 

Section  4.4

.    When  the  desired 

function  is displayed,  press  the 

SET

  button.    The  unit  will  return  to  the  last  menu  level.    To  exit  back  to  the  main  menu 

without changing the values, press the 

RECALL

 button.   

5.4.1

 

 



 

Set  Type  - 

Select  the  ALARM  TYPE.  Press  the 

SET

  button  to  enter  the  sub  menu  selections 

shown  below  and  use  the 

UP

 



  button  to  scroll  through  the  options.

 

Press  the 

SET

  button  to 

select the new setting or the 

RECALL

 button to exit without saving the settings.

 

5.4.1.1   



 

OFF - 

Turns the Alarm OFF so that it has no function.

 

5.4.1.2   



HIGH -

 Sets the Alarm to HIgh type. 

 

5.4.1.3   



LOW -

 Sets the Alarm to LOw type.

 

5.4.2

 

 



 

LATCH  -

  Press  the 

SET

  button  and  use  the 

UP

 



  button  to  select  whether  the  Alarm  is 

LATCHing  –  select 

YES

,  or  NON-LATCHING  –  select 

NO

.  Press  the 

SET

  button  to  select  the 

new setting or the 

RECALL

 button to exit without saving the settings.

 

5.4.3

 

 



 

LOCK -

 Press the 

SET

 button and use the 

UP

 



 button to select whether the Alarm will LOCk 

out – select 

YES

, or 

NO

. Press the 

SET

 button to select the new setting or the 

RECALL

 button to 

exit without saving the settings. 

Содержание ACT-3X

Страница 1: ...Totalizer Ratemeter User Manual and Reference Guide MONARCH INSTRUMENT 15 Columbia Drive Amherst NH 03031 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www mona...

Страница 2: ...e personal injury or death may occur as a result 7 This instrument may not be safe for use in certain hazardous environments and serious personal injury or death could occur as a result of improper us...

Страница 3: ...4 3 2 3 UP and DOWN BUTTONS 5 3 2 4 RECALL BUTTON 5 4 0 OPERATION 5 4 1 Modes of Operation Channel 1 5 4 1 1 RPM Mode 5 4 1 2 Frequency Mode 5 4 1 3 Scale Mode Ratemeter 5 4 1 4 Single Event Capture S...

Страница 4: ...5 9 11 5 10 11 5 11 12 5 12 12 6 0 SERIAL OUTPUT 12 6 1 Data from the Serial Interface 12 6 2 Serial Commands 12 6 2 1 Run Mode Commands 12 6 2 2 Control Commands 13 7 0 SPECIFICATIONS 14 8 0 OPTIONS...

Страница 5: ...xed equipment or permanently connected If permanently connected disconnection from the supply must be possible via a customer supplied switch or circuit breaker rated at 120V or 240V dependent on loca...

Страница 6: ...ve Hot wire to the terminal marked L and the Neutral Return wire to the terminal marked N NOTE The ground connection is not required as the unit is fully isolated from the mains If the unit is dc powe...

Страница 7: ...nput is also referred to as Channel 2 CH 2 and will accept a TTL input or any signal up to 12Vdc It has a weak pull up internally and can be used with a set of potential free contacts The AUX input is...

Страница 8: ...is another input pulse It is more useful at slow speeds as it appears to be on continuously at higher inputs It gives an indication that a valid input trigger signal is present 3 1 3 MAX and MIN LEDS...

Страница 9: ...1 2 Frequency Mode In the Frequency Mode the unit displays input pulses per second or more commonly Hertz Hz This is the most basic mode of operation The range of measurement in this mode is 0 0833 t...

Страница 10: ...nding on the sense programmed It acts as a gate preventing the input signal from updating the unit 4 2 3 Single Event Capture When programmed as Single Event Capture this input acts as the second inpu...

Страница 11: ...become de energized Thus in the event of a power failure the relays will drop out by default creating the equivalent of an alarm condition 4 5 Analog AO and Current IO Output The ACT 3X has options f...

Страница 12: ...or 14 640 RPM the internal update rate is the period of the input frequency Thus the response of the alarms etc can be seen to be a function of the input Above an input of 244 Hz the alarms respond wi...

Страница 13: ...d scale factor Press RECALL to exit up one level without making a change or press SET to save the change The edit buttons function as follows RESET changes the digit being edited blinking digit UP inc...

Страница 14: ...EGative or POSitive Press the SET button to select the new setting or the RECALL button to exit without saving the settings 5 3 Sets the decimal point on the displayed reading To enter this menu item...

Страница 15: ...a change or press SET to save the change 5 6 2 FULL SCALE Sets the Full Scale value for the Analog output Press the SET button and use the buttons to edit the value as per above Press RECALL to exit u...

Страница 16: ...ent occurs Other information is sent on demand and is covered in the following section The messages are sent as standard ASCII and all messages end with a carriage return CR There is no Line feed sent...

Страница 17: ...mand Category Argument Value eg CH_1 TYPE SCALE 60 0 sets the Scale Mode Scale factor to 60 0 Command Argument eg DECPT NONE sets the Decimal Point to none Command Category Argument Action CH_1 Channe...

Страница 18: ...100 times per second Standard Gate or up to 1000 Fast Gate Dependent on input frequency Input must be measurement rate Scale Mode Scaling Computation Programmable scaling 0 0001 to 99 999 Front panel...

Страница 19: ...preadsheet Software available for download at www monarchinstrument com ROLS W Remote Optical Laser Sensor with 8 foot cable ROS W Remote Optical Sensor with 8 foot cable ROS P 25 Remote Optical Senso...

Страница 20: ...ing the circumference of the shaft you are monitoring This could also be a speed wheel tire etc The circumference x diameter 3 14159 In order to scale we need to know what we want as opposed to what w...

Страница 21: ...il the display shows and then press the SET button The display will show the current scale factor value with the right most digit flashing The UP and DOWN buttons change the flashing digit The RESET b...

Страница 22: ...an error in the placement of the sensors The tape edges won t be exactly 1 inch apart Assuming the edges were really 1 01 inches apart there would be 1 error So at 130 MPH your reading would be 128 7...

Страница 23: ...hat the device is set up for DHCP by default so it will get an IP address from your local server Once the search is complete the window will be populated as shown next The current device setting is sh...

Страница 24: ...ons to set the required value Press RECALL to exit up one level without making a change or press SET to make the change The edit buttons function as follows RESET changes the digit being edited blinki...