Page 18

 

 
 

APPENDIX C 

– USING THE SINGLE EVENT CAPTURE MODE 

This is to how to calculate a scale factor and to show sources of measurement error. 

In this example, the distance between sensors is 1 inch and we want the readings displayed in Miles Per Hour (MPH).  The fastest 
measurement we intend to make is 130 MPH. 

First calculate the scale factor.  With a scale of one, the tachometer will display readings in pulses per second. 

The scale factor can be calculated as: 

1 Pulse 

1 Inch 

3600 Seconds 

  1 Foot 

   1 Mile 

0.05681818 Miles 

———   x  ———   x  ——————   x  ————   x  ————   =  ———————— 
Second 

 Pulse 

       Hour 

 12 Inches 

5280 Feet 

           Hour 

There will be an error because scale factors can only be five digits.  Therefore, the scale factor is rounded to 0.0568. 

You will also have an error in the placement of the sensors.  The tape edges won’t be exactly 1 inch apart.  Assuming the edges 
were really 1.01 inches apart there would be 1% error.  So at 130 MPH your reading would be 128.7 MPH. 

The internal clock inside the tachometer runs at 2 MHz.  All measurements are synchronized to this internal clock giving a ±0.5 
microsecond uncertainty.  As RPM, MPH, etc increase, the measurement time decreases. As the measurement time decreases, the 
small measurement uncertainty becomes a larger percentage of the measurement. 

At 130 MPH there would be 0.00043706293706 seconds between pulses. (Seconds  = scale factor / 130 MPH.)  This is equal to 
about 874 clock cycles for a 2 MHz internal reference clock. 

873 clock cycles = 130.16766 MPH 
875 clock cycles = 129.87013 MPH 

Therefore, we have an error of ±0.148765 MPH from the clock resolution at 130 MPH.  The sensors must be placed further apart 
for better resolution. 

Содержание ACT-3X

Страница 1: ...Totalizer Ratemeter User Manual and Reference Guide MONARCH INSTRUMENT 15 Columbia Drive Amherst NH 03031 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com Website www mona...

Страница 2: ...e personal injury or death may occur as a result 7 This instrument may not be safe for use in certain hazardous environments and serious personal injury or death could occur as a result of improper us...

Страница 3: ...4 3 2 3 UP and DOWN BUTTONS 5 3 2 4 RECALL BUTTON 5 4 0 OPERATION 5 4 1 Modes of Operation Channel 1 5 4 1 1 RPM Mode 5 4 1 2 Frequency Mode 5 4 1 3 Scale Mode Ratemeter 5 4 1 4 Single Event Capture S...

Страница 4: ...5 9 11 5 10 11 5 11 12 5 12 12 6 0 SERIAL OUTPUT 12 6 1 Data from the Serial Interface 12 6 2 Serial Commands 12 6 2 1 Run Mode Commands 12 6 2 2 Control Commands 13 7 0 SPECIFICATIONS 14 8 0 OPTIONS...

Страница 5: ...xed equipment or permanently connected If permanently connected disconnection from the supply must be possible via a customer supplied switch or circuit breaker rated at 120V or 240V dependent on loca...

Страница 6: ...ve Hot wire to the terminal marked L and the Neutral Return wire to the terminal marked N NOTE The ground connection is not required as the unit is fully isolated from the mains If the unit is dc powe...

Страница 7: ...nput is also referred to as Channel 2 CH 2 and will accept a TTL input or any signal up to 12Vdc It has a weak pull up internally and can be used with a set of potential free contacts The AUX input is...

Страница 8: ...is another input pulse It is more useful at slow speeds as it appears to be on continuously at higher inputs It gives an indication that a valid input trigger signal is present 3 1 3 MAX and MIN LEDS...

Страница 9: ...1 2 Frequency Mode In the Frequency Mode the unit displays input pulses per second or more commonly Hertz Hz This is the most basic mode of operation The range of measurement in this mode is 0 0833 t...

Страница 10: ...nding on the sense programmed It acts as a gate preventing the input signal from updating the unit 4 2 3 Single Event Capture When programmed as Single Event Capture this input acts as the second inpu...

Страница 11: ...become de energized Thus in the event of a power failure the relays will drop out by default creating the equivalent of an alarm condition 4 5 Analog AO and Current IO Output The ACT 3X has options f...

Страница 12: ...or 14 640 RPM the internal update rate is the period of the input frequency Thus the response of the alarms etc can be seen to be a function of the input Above an input of 244 Hz the alarms respond wi...

Страница 13: ...d scale factor Press RECALL to exit up one level without making a change or press SET to save the change The edit buttons function as follows RESET changes the digit being edited blinking digit UP inc...

Страница 14: ...EGative or POSitive Press the SET button to select the new setting or the RECALL button to exit without saving the settings 5 3 Sets the decimal point on the displayed reading To enter this menu item...

Страница 15: ...a change or press SET to save the change 5 6 2 FULL SCALE Sets the Full Scale value for the Analog output Press the SET button and use the buttons to edit the value as per above Press RECALL to exit u...

Страница 16: ...ent occurs Other information is sent on demand and is covered in the following section The messages are sent as standard ASCII and all messages end with a carriage return CR There is no Line feed sent...

Страница 17: ...mand Category Argument Value eg CH_1 TYPE SCALE 60 0 sets the Scale Mode Scale factor to 60 0 Command Argument eg DECPT NONE sets the Decimal Point to none Command Category Argument Action CH_1 Channe...

Страница 18: ...100 times per second Standard Gate or up to 1000 Fast Gate Dependent on input frequency Input must be measurement rate Scale Mode Scaling Computation Programmable scaling 0 0001 to 99 999 Front panel...

Страница 19: ...preadsheet Software available for download at www monarchinstrument com ROLS W Remote Optical Laser Sensor with 8 foot cable ROS W Remote Optical Sensor with 8 foot cable ROS P 25 Remote Optical Senso...

Страница 20: ...ing the circumference of the shaft you are monitoring This could also be a speed wheel tire etc The circumference x diameter 3 14159 In order to scale we need to know what we want as opposed to what w...

Страница 21: ...il the display shows and then press the SET button The display will show the current scale factor value with the right most digit flashing The UP and DOWN buttons change the flashing digit The RESET b...

Страница 22: ...an error in the placement of the sensors The tape edges won t be exactly 1 inch apart Assuming the edges were really 1 01 inches apart there would be 1 error So at 130 MPH your reading would be 128 7...

Страница 23: ...hat the device is set up for DHCP by default so it will get an IP address from your local server Once the search is complete the window will be populated as shown next The current device setting is sh...

Страница 24: ...ons to set the required value Press RECALL to exit up one level without making a change or press SET to make the change The edit buttons function as follows RESET changes the digit being edited blinki...

Отзывы: